INSTALLATION OF MEDIUM VACUUM 
PUMP DOWNSTREAM INTERCONNECT KIT 

The above figure shows the basic interconnection 
plumbing when using the Medium Flow Vacuum Pump 
with the molbloc-S.  This kit includes 2 conical reducers 
with the ISO-KF Style Vacuum Flange ends. When 
connecting a single molbloc-S to the vacuum pump, 
use both the 40 mm x 25 mm, and the 25 mm x 16 mm 
conical reducers.  When utilizing the downstream tee 
assembly (

DHI 

P\N 401884) only the 40 mm x 25 mm 

conical reducer is required. 

For the sake of flexibility this interconnection kit 
includes a coil of 1 1/2 in. wire reinforced PVC hose.  
This hose can be cut to the precise length required in 
each application.  Hose barb fittings slip inside the 
hose and are secured by worm drive clamps. 

VACUUM SHUT-OFF VALVE 

The vacuum shut-off valve is provided so that the 
operator can continue to run the vacuum pump while 
changing the Device Under Test (DUT) or the molbloc-S 
between calibration runs.  It is not recommended to 
start and stop the pump more than a few times per day.  
Therefore continue to run the pump with the gas ballast 
valve open and the vacuum shut-off valve closed.  
Power consumption is minimal when the pump is 
running at its ultimate pressure. 

CONNECTING, DISCONNECTING AND 
MOUNTING MOLBLOC-S WITH ISO-KF 
STYLE VACUUM FLANGES (MOLBLOC-S 
OUTLET FITTING) 

ISO-KF Style Vacuum Flange Fittings 

The outlet connection system on the molbloc-S body is 
the ISO-KF style vacuum flange.  It utilizes an internal 
centering ring, overpressure ring, and an external clamp.  
The seal is effected by the uniform application of 
pressure by the clamp on the 15° surface of the mating  

stainless steel flanges.  These mating flange surfaces 
compress a Viton

®

 O-ring that is held in place by the 

centering ring.  The overpressure ring keeps the O-ring 
in place, and maintains a leak free connection when 
the system is subjected to internal pressures above 
vacuum level.  This connection is reusable, rotatable, 
and can operate leak free in vacuum applications up to 
10

-8

 Torr.  When an overpressure ring is used, KF25 or 

smaller flange connections can operate leak free to 
positive pressure beyond 600 kPa (87 psia) 

Making and Breaking Procedure for ISO-KF 
Style Vacuum Flange Connections 

  Place the centering ring (and overpressure ring if 

applicable) into the groove of one of the ISO-KF 
flanges to be mated. 

  Align the mating flange, and close the gap by hand.  

Hold the flanges together while putting the clamp 
around them. 

  Swing the clamp closed.  If necessary loosen the 

wing nut to allow the thrust washer and wing nut to 
fall into position on top of the clamp.  Fully tighten 
the wing nut by hand. 

  To break the fitting, hold one of the plumbing 

sections with your hand and loosen the wing nut 
until the clamp can be separated.  Remove the clamp, 
centering ring, (and overpressure ring if applicable). 

 

 