Dewert User Manual 2000XL Series– Updated 05.2021/1 MDR Page 12 of 19
2006XLE
2006XL/H
2008XLE
2008XL/H
Models
2007XLE
2007XL/H
2009XLE
2009XL/H
Max. length (mm)
1950/1550*
1950/1550*
2190/1790*
2190/1790*
Width (mm)
700; 800
700; 800
700; 800
700; 800
Head and foot part length (mm)
450
450
450
450
Weight (approx., depending on
equipment) kg
75
75
80
80
Min.
– max. height (mm)
520 - 960
540 - 960
520 - 960
540 - 960
Adjustment time (motor) (s)
22
-
22
-
Head part adjustment range (gas spring)
-90° / +30°
-90° / +30°
-40° / +55°
-40° / + 55°
Foot part adjustment range
-
-
-90° / +30°
-90° / +30°
Max. patient weight (kg)
250***
250***
250***
250***
2010XLE
2010XL/H
2114XLE
2114XL/H
Models
2011XLE
2011XL/H
2115XLE
2115XL/H
Max. length (mm)
1950/1550*
1950/1550*
1400
1400
Width (mm)
700; 800
700; 800
650; 800
650; 800
Head and foot part length (mm)
450
450
-
-
Weight (approx., depending on
equipment) kg
75
75
75
75
Min.
– max. height (mm)
530 - 970
550 - 970
480 - 920
500 - 920
Adjustment time (motor) (s)
22
-
22
-
Head part adjustment range (gas spring)
0° / +60°
0° / +60°
-
-
Foot part adjustment range
-90° / + 30°
-90° / +30°
-
-
Max. patient weight (kg)
250***
250***
250
250
2030XLE
2030XL/H
Models
2035XLE
2035XL/H
Max. length (mm)
1950
1950
Width (mm)
700; 800
700; 800
Head part length (mm)
750
750
Weight (approx., depending on
equipment) kg
75; 80
75; 80
Min.
– max. height (mm)
480 - 920
500 - 920
Adjustment time (motor) (s)
22
-
Head part adjustment range (gas spring)
-25° / +70°
-25° / +70°
Max. patient weight (kg)
250
250
The indicated table height may vary depending on the equipment, e.g.
castors with 100 mm diameter - individually or centrally lockable - height: +20 mm
With other castor sizes, depending on the castor diameter
*
Foot/head part in horizontal/vertical position
**
With version -01 (chrome plated upper frame) height adjustment range: +20 mm
***
Weight load not on head and foot part, only centrally on the fixed lying surface
The motor is equipped with a thermal protective switch as standard. This causes the motor to cut out if the weight
load is too extreme or the motor’s duty cycle (DC: 25 s/400 s) is exceeded. The table should be ready for use again
after a 15-minute rest period. There is, therefore, no risk of overloading the motor.
The mechanics of the tables have been designed with extensive safety margins in mind. The max. patient weight
is provided with a 4-fold static safety factor, i.e. the design has been fully tested for a 4-fold load.
Technical data subject to change.