TROUBLESHOOTING
DANGER
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable internal components. If service is required, return unit to a
qualified DeVilbiss Healthcare provider or an authorized service center. Opening or tampering with the unit will void warranty.
NOTE
– Your DeVilbiss Suction Unit contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO your DeVilbiss
Healthcare PROVIDER WHERE YOU PURCHASED IT, please take a few moments to check for these possible causes:
PROBLEM
ACTION
Unit does not turn on when external power is connected. Green
external power light does not illuminate.
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
Pump runs, but there is no suction.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks or leaks.
3. Ensure that float shut-off in collection container is not activated or filter cartridge occluded.
4. Check for leaks or cracks in collection container assembly.
Low suction.
1. Use vacuum regulator knob to increase suction level.
2. Check system for leaks.
Unit does not turn on (no external power is connected). 7314P Series
Only
1. Check that battery is fully charged and/or charge battery.
Battery will not charge (external power and charge indicator lights
should be illuminated during charge mode) 7314P Series Only
1. Verify that both external power and charge indicator lights illuminate.
2. Check power sources and connections.
3. Ensure wall outlet is live by plugging in a lamp.
SPECIFICATIONS/CLASSIFICATIONS
Size (including container)
8.3 H x 8.0 W x 8.5 D (21.1 cm x 20.3 cm x 21.6 cm) (not including AC to DC adapter)
Weight (including container)
7314P Series - 6.6 lb. (3 kg) (not including AC to DC adapter)
7314D Series - 4.3 lb. (2.0 kg)
Typical Operating Sound Level
55 dBA
Electrical Requirements
100-240V~, 50/60Hz, 1.2A max
; 12V ; 33 W Max
Vacuum Range
50-550 mm Hg +/- 10%*
Air Flow @ Pump Inlet
27 LPM (free flow) typical (may be less when running from internal battery)* (NOTE– Using Flovac Liners may impair the
device’s performance.)
Disposable Collection Container Capacity
800 ml (cc)
Reusable Collection Container Capacity
1200 ml (cc)
Warranty
Two-years limited, excluding internal battery (7314P series only) and collection container
Internal Battery (7314P Series Only)
90-day
Approvals
IEC 60601-1:2005+AMD1:2012; EN 60601-1-2: 2015; ISO 10079-1:2015; IEC 60601-1-2:2014; IEC 60529:1989+AMD1:1999
+AMD2:2013; IEC 60601-1-6:2010; IEC 60601-1-11:2015;IP22
CONFORMS TO AAMI STD. ES 60601-1, AAMI STD. HA 60601-1-11, IEC STD. 60601-1-6, ISO STD. 10079-1
CERTIFIED TO CSA STD. C22.2 NO. 60601-1, NO. 60601-1-11, IEC STD. 60601-1-6, ISO STD. 10079-1
7314 Series complies with IEC 60601-1 3rd Edition
Environmental Conditions
Operating Temperature Range
32°F (0°C) - 104°F (40°C)
Operating Relative Humidity
0-95%
Operating Atmospheric Pressure
10.2 psi (70 kPa) - 15.4 psi (106 kPa)
Storage & Transport Temperature Range
-40°F (-40°C) - 158°F (70°C)
Storage & Transport Relative Humidity
0-95%
Storage & Transport Atmospheric Pressure
7.3 psi (50 kPa) - 15.4 psi (106 kPa)
Equipment Classifications
With respect to protection from electric shock
Class II and internally powered
Degree of protection against electric shock
Type BF Applied Parts
Degree of protection against ingress of liquids
IP22 and ordinary power supply
Mode of Operation
Intermittent Operation: 30 minutes on, 30 minutes off
Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide.
ISO Classification
High Flow/High Vacuum
7314 series - Electrically powered medical suction equipment for non-transport use according to EN ISO 10079-1:2015
* Conditions may vary based on altitude above sea level, barometric pressure, and temperature.