System Specifications
Classification
Class II, internally powered Type BF equipment
Applied Part
Probes (MEP10, MEP13)
Environmental Conditions
Transportation and Storage:
•
-20°C to +50°C (-4°F to 122°F)
•
10% to 95% Relative Humidity, Non-Condensing
•
500 to 1060 hPa Atmosphere Pressure
Operation:
•
+5°C to +40°C (41°F to 104°F)
•
30% to 75% Relative Humidity, Non-Condensing
•
810 to 1060 hPa Atmosphere Pressure
Mode of Operation
Non-Continuous: 40 cycles ON, 30 minutes OFF
Holster Package Specifications
Package Materials
B-Flute Corrugate Carton
EPE Foam
Environmental Conditions
Fully Loaded carton with product: 3lb 1.5oz:
•
EPE Foam: 5.7oz
•
B-Flute Corrugate Carton with Label: 10.4oz
•
Paper IFU + Set Up Sheet: 4.0oz
•
Charger: 14.1oz
•
Holster: 8.6oz
•
A Box: 7.6oz
Electromagnetic Compatibility (EMC), Guidance, and Manufacturer
Declarations
The Mammotome elite Biopsy System requires special precautions regarding
electromagnetic compatibility (EMC) and must be installed and used in accordance with
the EMC information provided in this installation guide. The Mammotome elite Biopsy
System is intended for use in the electromagnetic environment specified below. The
customer or user of the Mammotome elite Biopsy System should assure that is used in
such an environment.