![Devicor Mammotome elite Instructions Manual Download Page 15](http://html1.mh-extra.com/html/devicor/mammotome-elite/mammotome-elite_instructions-manual_2488821015.webp)
14
including tools and solutions, may influence the wear of a device or equipment.
There are no limitations for time of storage of any disinfected device prior to use.
Additionally, no support systems are required.
Cleaning the Holster
Less than 45 minutes after each use, manually clean the Holster thoroughly utilizing a
pH neutral enzymatic detergent. Proceed with manual cleaning utilizing an enzymatic
detergent as follows:
1. Prepare a pH neutral detergent or pH neutral enzymatic detergent according to
the detergent manufacturer’s directions.
2. Use a soft, clean cloth lightly moistened with the cleaning solution to manually
clean all surfaces that are exposed when the probe is attached to the Holster. In
addition, use a standard soft bristle toothbrush to clean areas containing
crevices. The gears and electronics on the underside of the Holster should not be
wiped with detergent.
3. To rinse, wipe thoroughly using a soft, clean cloth lightly moistened with
lukewarm (27°C – 44°C) tap water.
4. To rinse the crevices, wipe thoroughly using a clean soft bristle, standard
toothbrush moistened with lukewarm (27˚C
-
44˚C) tap water.
5. Dry with a clean, soft cloth. Do not expose to heat greater than 50°C
6. Inspect the device for visible contamination. If contamination is present, repeat
steps 1-5 until no visible contamination is present.
CAUTION
: Do not attempt to sterilize the Holster or Charger Base through autoclave,
ethylene oxide, radiation, or plasma sterilization procedures; do not process through an
automated washer-disinfect or ultrasound bath; do not spray the Holster with fluids or
submerge in fluids; do not use liquids on the Charger Base; this may damage the
instrument. Do not allow the detergent to come in contact with the electronic
components on the underside of the Holster as this may damage the Holster. If the
Holster is cleaned improperly, the warranty may be void. Do not transport the device in
a container prior to cleaning.
Disinfecting the Holster
An intermediate level disinfection process must follow the cleaning procedure. The
following chemical disinfectants are approved for use with the Holster:
•
50% Bleach (Sodium Hypochlorite) Solution (1 part bleach to 1 part purified or
deionized water)