3. The Holster is not charging when placed into the Charging Base:
Check that the AC power cord is securely inserted into the Charging Base and a
wall outlet. Check that there are no obstructions over the charging connections
on the Charging Base or Holster and then insert the Holster into the Charging
Base. Ensure the audible tone sounds from the Holster and the Battery Indicator
Light flashes. If there is no audible tone or flashing light, replace the Charging
Base and AC power cord by contacting the MAMMOTOME Customer Support
Center or a representative.
Holster Specifications
Electrical Conformity
This medical equipment has passed all required testing for electric
shock, fire and mechanical hazards in accordance with UL 60601-1,
IEC/EN 60601-1, CAN/CSA C22.2 No 60601-1
Fluid Ingress
Not protected against the ingress of fluids
Flammability
Equipment not suitable for use in the presence of flammable
anesthetic
Dimensions
Length: 18.9 cm
Width: 5.0 cm
Height: 3.8 cm
Battery Location
Batteries are located in the proximal end of the Holster
AC Power Cord and Adapters
Use only with supplied AC power accessories:
•
Model : FRM015-S09-1 p/n TRM15-S10-I-351359
•
Input Voltage: 100-240VAC
•
Input Current: 0.7A-0.4A
•
Input Frequency: 50-60Hz
•
Output Voltage:10 Volts DC
•
Output Current: 1.5 A
•
Max Watts: 15W