The “current status” field .
●
The “Battery: X.XX V” states the current Voltage of the battery. This parameter must
report at least 5.00 volts value. If this value shows less than 5.00 Volts STOP! DO NOT
deploy the ARC unit. This is an indication that the batteries must be changed.
●
The “Release: XX.XX V” states the Voltage necessary to trigger a release mechanism.
This parameter must report at least 25.00 Volts value. If this value shows less than 25.00
Volts, STOP. DO NOT deploy the ARC unit. This is an indication that the battery must be
changed.
If you need to change the batteries, please read the section of this manual guiding you on how to
install/change the batteries of an ARC unit. Once the new batteries are installed, repeat the
communications test.
●
If the “Battery” Voltage reported at 5.00 Volts or more, AND the “Release” voltage
reported at 25.00 Volts or more, this means battery charge levels are acceptable and the
ARC unit is ready to be deployed underwater.
●
This concludes the communication test between the ARC submergible unit and the STM
surface station. The ARC unit is now listed as registered in the “Acoustic Release