Desco SCS-73627 Installation, Operation And Maintenance Manual Download Page 6

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Resistivity and Resistance Per ANSI/ESD STM11.11 
section 12.0 

CONVERSION TO RESISTIVITY, 

When it is appropriate to convert a resistance 
obtained by this test method to an equivalent 
resistivity in ohms per square, multiply the resistance 
measurements obtained by this method by ten. The 
conversion factor of ten is derived from the geometry 
of the electrode assembly. NOTE: Conversions to 
resistivity may not be valid for materials that are 
laminated, plated or metallized with conductive 
materials. This fact should be understood when 
users of this test procedure are required to satisfy 
specifications given in 

ohms/square.

 

 

Measure Samll Samples or Small Area on a Large 
Sample 

1.  Place the sample to be tested on an insulative 

surface.   

2.  Connect both test leads to the Two-Point-Probe 
3.  Compress the spring-loaded pins on the 

Two-Point-Probe downward onto the sample to 
the end. Ensure that your skin does not make 
contact with any of the metal on the probe 

4.  Push the red pushbutton to perform a test. 

Should the 15-second electrification period 
appear, it may be bypassed by pushing the red 
pushbutton a second time. 

5.  Push the right black pushbutton to save the 

measurement if desired. 

 

Measure Floor Materials and Footwear- 
Resistance Measurement in Combination with a 
Person (ANSI/ESD STM97.1) 

1.  Connect the red test lead to the Handheld- 

Electrode, connect the black lead to the Single- 
Surface-Electrode. 

2.  Put the Single-Surface-Electrode on the floor, a 

person stands on the floor and hold the Handheld-
Electrode. 

3.  Push the red pushbutton to perform a test. 

Should the 15-second electrification period 
appear, it may be bypassed by pushing the red 
pushbutton a second time. 

4.  Push the right black pushbutton to save the 

measurement if desired. 

 

SCS 

- 926 JR Industrial Drive, Sanford, NC 27332 

East: (919) 718-0000 | West: (909) 627-9634 

 Website: 

StaticControl.com 

TB-9092 

Page 6 of 8 December 2019                                                                                                     

©

 2019 DESCO INDUSTRIES INC 

Employee  Owned

Maintenance 

The Resistance Pro Meter requires little 
maintenance. There are no user serviceable parts. 
If the meter requires service beyond cleaning the 
electrodes or replacing the batteries. 
 

Battery Replacement 

Replace the batteries once the battery life indictator 
is empty. Open the compartment located on the 
back of the meter to replace the batteries. The 
meter uses four AA alkaline batteries. Ensure that 
the batteries

 polarities are oriented in the correct 

fashion to avoid any possible circuit damage. 

 
Cleaning the Resistance Pro Meter 

The area surrounding the test jacks at the top end 
of the meter should be wiped with a clean, 
isopropanol-alcohol moistened cloth to remove skin 
oils that will accumulate and affect the meter

accuracy at high resistances. The frequency of 
cleaning will depend on usage. SCS recommends 
cleaning this area once a month. Cable jackets 
should also be cleaned in this fashion. 
 

Cleaning the Electrodes 

Per ANSI/ESD S4.1 

Clean the electrodes with a 

minimum 70% isopropanol-water solution. Make 
sure the electrodes

 conductive pads are dry prior to 

use.

   

 
See specific product test standards for test lab 
specimen cleaning instructions. Per ANSI/ESD S4.1 
Worksurfaces 

The test specimens and electrodes 

shall be cleaned twice with a minimum 70% 
isopropanol-water solution using a clean, low-linting 
cloth each time.

 (Note: The item should then be 

conditioned for 72 hours minimum) 

Summary of Contents for SCS-73627

Page 1: ...s ANSI ESD S7 1 Resistive Characterization of Materials Floor Materials l Foot Grounders ESD SP9 2 Foot Grounders Resistive Characterization l Garments ANSI ESD STM2 1 Garments Seating ANSI ESD STM12 1 Seating Resistive Measurement l Floor Footwear ANSI ESD STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person A Compliance Verification Plan shall be established t...

Page 2: ...o Meter Kit 1 Resistance Pro Meter 2 Test Leads 5 Length 1 Single Surface Electrode 1 Dual Surface Electrode 4 AA Alkaline Batteries 1 Ground Plug Adapter 1 Gator Clip 1 Plastic Carrying Case 1 Certificate of Calibration SCS 926 JR Industrial Drive Sanford NC 27332 East 919 718 0000 West 909 627 9634 Website StaticControl com TB 9092 Page 2 of 8 December 2019 2019 DESCO INDUSTRIES INC Employee Own...

Page 3: ...8 10 ohms or 100 000 000 ohms C OLED Display Displays the temperature relative humidity battery life test voltage and resistance measurement D Power Switch Slide the switch to the left to power the meter OFF Slide the switch to the right to power the meter ON E Black Pushbuttons Each black pushbutton corresponds to the prompts on the bottom left and bottom right of the display These buttons are us...

Page 4: ...te StaticControl com TB 9092 Page 4 of 8 December 2019 2019 DESCO INDUSTRIES INC Employee Owned Auto Enables a 15 second electrification period when the measured resistance is 1 x 1010 ohms or greater to maintain test accuracy Fixed Complies with ANSI ESD S4 1 and enables A 15 second electrification period when the measured resistance is 1 x 106 ohms or greater P Beep Enables and disables the audi...

Page 5: ...ments by placing the electrode on the most commonly used or worn area SCS 926 JR Industrial Drive Sanford NC 27332 East 919 718 0000 West 909 627 9634 Website StaticControl com TB 9092 Page 5 of 8 December 2019 2019 DESCO INDUSTRIES INC Employee Owned Measure Resistance Point to Point Rtt on the Surface 1 Do not clean the surface 2 Remove from the surface only those items that might interfere with...

Page 6: ... the 15 second electrification period appear it may be bypassed by pushing the red pushbutton a second time 4 Push the right black pushbutton to save the measurement if desired SCS 926 JR Industrial Drive Sanford NC 27332 East 919 718 0000 West 909 627 9634 Website StaticControl com TB 9092 Page 6 of 8 December 2019 2019 DESCO INDUSTRIES INC Employee Owned Maintenance The Resistance Pro Meter requ...

Page 7: ...adings The meter cannot be normalized inside objects enclosed boxes containers or cases The temperature inside an enclosed case will differ from the outside temperature SCS 926 JR Industrial Drive Sanford NC 27332 East 919 718 0000 West 909 627 9634 Website StaticControl com TB 9092 Page 7 of 8 December 2019 2019 DESCO INDUSTRIES INC Employee Owned These cases will act as insulators The meter must...

Page 8: ...ements Operating Temperature 41ºF to 85ºF 5ºC to 30ºC Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humidity of 80 up to 85 F 30 C Dimensions meter 3 94 x 8 27 x 1 26 100mm x 210mm x 32 mm Dimensions carrying case 15 0 x 14 5 x 3 5 381mm x 368mm x 89mm Weight meter with batteries 0 9 lbs 0 4 kg Weight populated carrying case 13 0 lbs 5 9 kg Country...

Reviews: