background image

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Resistance Pro Meter Kit 
Installation, Operation and Maintenance

 

 

Figure 1. SCS-73627 Resistance Pro Meter Kit

 

 

Description 

The SCS-73627 Resistance Pro Meter Kit is   
an instrument designed to measure resistance 
point-to-point (Rtt) or surface to ground (Rtg). 
 
Its test functions include: 

l

 

Resistance measurement accuracy of 

±

10% 

(

±

20% accuracy for 1 x 10

12

 ohms and greater) 

l

 

Resistance range of <1 x 10

3

 ohms to >1 x 10

12

 

ohms 

l

 

Under load voltages of 10 and 100 volts 

±

5% 

l

 

Electrification period of approximately 15 seconds 

 
The Resistance Pro Meter also measures ambient 
temperature and relative humidity. Up to 100 
measurements may be stored and recalled from the 
meter's internal memory. This includes the resistance 
value, temperature, relative humidity and test voltage 
at the time of the measurement. 

SCS 

- 926 JR Industrial Drive, Sanford, NC 27332 

East: (919) 718-0000 | West: (909) 627-9634 

 Website: 

StaticControl.com 

TB-9092 

Page 1 of 8 December 2019                                                                                                     

©

 2019 DESCO INDUSTRIES INC 

Employee  Owned

The Surface Resistance Meter is referenced and 
designed to be used to make measurements in 
accordance with the test procedures in: 
 

l

 

Compliance Verification-ESD TR53-Resistance 
Measurements 

l

 

Worksurfaces-ANSI/ESD S4.1 Worksurfaces 

l

 

Floors-ANSI/ESD S7.1-Resistive- 
Characterization of Materials Floor Materials 

l

 

Foot Grounders-ESD SP9.2-Foot Grounders 
Resistive Characterization 

l

 

Garments-ANSI/ESD STM2.1 Garments 
Seating-ANSI/ESD STM12.1-Seating-Resistive 
Measurement 

l

 

Floor/Footwear-ANSI/ESD STM97.1-Floor 
Materials and Footwear- Resistance 
Measurement in Combination with a Person 

 

A Compliance Verification Plan shall be established 

to ensure the Organization

s fulfillment of the 

technical requirements of the ESD Control Program 
Plan. Process monitoring (measurements) shall be 
conducted in accordance with a Compliance 
Verification Plan that identifies the technical 
requirements to be verified, the measurement limits 
and the frequency at which those verifications shall 
occur. The Compliance Verification Plan shall 
document the test methods and equipment used for 
process monitoring and measurements. If the test 
methods used by the Organization differ from any of 
the standards referenced in this document, then 
there must be a tailoring statement that is 
documented as part of the ESD Control Program 
Plan. Compliance verification records shall be 
established and maintained to provide evidence of 
conformity to the technical requirements. The test 
equipment selected shall be capable of making the 
measurements defined in the Compliance 
Verification Plan.

 (ANSI/ESD S20.20 section 7.3)

 

SCS 

USER GUIDE TB-9092 

 

Made in the 

United States of America 

Summary of Contents for SCS-73627

Page 1: ...s ANSI ESD S7 1 Resistive Characterization of Materials Floor Materials l Foot Grounders ESD SP9 2 Foot Grounders Resistive Characterization l Garments ANSI ESD STM2 1 Garments Seating ANSI ESD STM12 1 Seating Resistive Measurement l Floor Footwear ANSI ESD STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person A Compliance Verification Plan shall be established t...

Page 2: ...o Meter Kit 1 Resistance Pro Meter 2 Test Leads 5 Length 1 Single Surface Electrode 1 Dual Surface Electrode 4 AA Alkaline Batteries 1 Ground Plug Adapter 1 Gator Clip 1 Plastic Carrying Case 1 Certificate of Calibration SCS 926 JR Industrial Drive Sanford NC 27332 East 919 718 0000 West 909 627 9634 Website StaticControl com TB 9092 Page 2 of 8 December 2019 2019 DESCO INDUSTRIES INC Employee Own...

Page 3: ...8 10 ohms or 100 000 000 ohms C OLED Display Displays the temperature relative humidity battery life test voltage and resistance measurement D Power Switch Slide the switch to the left to power the meter OFF Slide the switch to the right to power the meter ON E Black Pushbuttons Each black pushbutton corresponds to the prompts on the bottom left and bottom right of the display These buttons are us...

Page 4: ...te StaticControl com TB 9092 Page 4 of 8 December 2019 2019 DESCO INDUSTRIES INC Employee Owned Auto Enables a 15 second electrification period when the measured resistance is 1 x 1010 ohms or greater to maintain test accuracy Fixed Complies with ANSI ESD S4 1 and enables A 15 second electrification period when the measured resistance is 1 x 106 ohms or greater P Beep Enables and disables the audi...

Page 5: ...ments by placing the electrode on the most commonly used or worn area SCS 926 JR Industrial Drive Sanford NC 27332 East 919 718 0000 West 909 627 9634 Website StaticControl com TB 9092 Page 5 of 8 December 2019 2019 DESCO INDUSTRIES INC Employee Owned Measure Resistance Point to Point Rtt on the Surface 1 Do not clean the surface 2 Remove from the surface only those items that might interfere with...

Page 6: ... the 15 second electrification period appear it may be bypassed by pushing the red pushbutton a second time 4 Push the right black pushbutton to save the measurement if desired SCS 926 JR Industrial Drive Sanford NC 27332 East 919 718 0000 West 909 627 9634 Website StaticControl com TB 9092 Page 6 of 8 December 2019 2019 DESCO INDUSTRIES INC Employee Owned Maintenance The Resistance Pro Meter requ...

Page 7: ...adings The meter cannot be normalized inside objects enclosed boxes containers or cases The temperature inside an enclosed case will differ from the outside temperature SCS 926 JR Industrial Drive Sanford NC 27332 East 919 718 0000 West 909 627 9634 Website StaticControl com TB 9092 Page 7 of 8 December 2019 2019 DESCO INDUSTRIES INC Employee Owned These cases will act as insulators The meter must...

Page 8: ...ements Operating Temperature 41ºF to 85ºF 5ºC to 30ºC Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humidity of 80 up to 85 F 30 C Dimensions meter 3 94 x 8 27 x 1 26 100mm x 210mm x 32 mm Dimensions carrying case 15 0 x 14 5 x 3 5 381mm x 368mm x 89mm Weight meter with batteries 0 9 lbs 0 4 kg Weight populated carrying case 13 0 lbs 5 9 kg Country...

Reviews: