Installation

TESTER CONFIGURATION

The resistance limits for footwear and wrist strap tests 

are controlled by the DIP switches located on the left 

side of the Combo Tester X3. Use the following tables 

for the DIP switch settings and their corresponding test 

values.

Footwear Resistance

DIP switches 1 and 2 control the HIGH test limit.

Switch 1 Switch 2 HIGH Limit Resistance

ON

ON

10 Megohms (1 x 10

7

)

OFF

OFF

35 Megohms (3.5 x 10

7

)

ON

OFF

100 Megohms (1 x 10

8

)

OFF

ON

1 Gigohm (1 x 10

9

)

DIP switches 3 and 4 control the LOW test limit.

Switch 3 Switch 4 LOW Limit Resistance

OFF

OFF

footwear test disabled

ON

OFF

100 Kilohms (1 x 10

5

)

OFF

ON

750 Kilohms (7.5 x 10

5

)

default setting

NOTE: At 1 Gigohm high limit resistance, a dirty foot 

plate could result in a false pass. Be sure to keep the 

foot plate clean particularly when using this setting. This 

setting is not suitable for relative humidity greater than 

50%.

Wrist Strap Resistance

DIP switches 5 and 6 control the HIGH test limit.

Switch 5 Switch 6 HIGH Limit Resistance

OFF

OFF

wrist strap test disabled

ON

ON

10 Megohms (1 x 10

7

)

ON

OFF

35 Megohms (3.5 x 10

7

)

default setting

DIP switch 5 must be ON (default setting) for the wrist 

strap test to be active. The wrist strap test will be 

disabled if DIP switch 5 is set to OFF.

The LOW limit for the wrist strap test is set to 750 

kilohms and cannot be modified by the user.

5.  Insert the ground cord’s banana plug into the 

ground jack located at the bottom of the tester. 

Connect ground cord’s ring terminal to equipment 

ground. This connection will remove any static 

charge from the user before the test.

 NOTE: Failure 

to correctly ground the Combo Tester X3 may result 

in damage not covered under warranty.

 

6.  Power the Combo Tester X3 using the included 

power supply.

Figure 3. Routing the foot plate cord through the foot 

plate’s guide

Bottom Side

INSTALLING THE COMBO TESTER X3

1.  Mount the tester at the desired location using the 

four mounting holes located in the corners of the 

yellow mounting plate. 

2.  Set the Dual Foot Plate below the tester. 
3.  Insert one end of the foot plate cord into the stereo 

jack located at the bottom of the tester. Insert the 

opposite end of the cord into the stereo jack located 

at the back of the foot plate. 

4.  Use the guide located on the bottom of the foot 

plate to route the cord out of the side. This will 

prevent the foot plate cord from being accidentally 

tripped and unplugged. 

Summary of Contents for Combo Tester X3

Page 1: ...r Combo Tester X3 50755 ESD Glove Test Fixture 50424 Limit Comparator 19262 Power Adapter 100 240VAC Input 12VDC 0 5A Output All Plugs 50406 Replacement Dual Foot Plate Per ANSI ESD S20 20 section 7 3 A Compliance Verification Plan shall be established to ensure the Organization s fulfillment of the technical requirements of the ESD Control Program Plan Process monitoring measurements shall be con...

Page 2: ...Switch Use this DIP switch to configure the resistance limits of the tester See the section titled Tester Configuration for more information B Footwear Status LEDs Displays the footwear test results C 12VDC Power Jack Connect the included power adapter here to power the Combo Tester X3 D Steady State Test Switch Place and hold your finger here to begin the test E Single Wire Wrist Strap Jack Inser...

Page 3: ...it Resistance OFF OFF wrist strap test disabled ON ON 10 Megohms 1 x 107 ON OFF 35 Megohms 3 5 x 107 default setting DIP switch 5 must be ON default setting for the wrist strap test to be active The wrist strap test will be disabled if DIP switch 5 is set to OFF The LOW limit for the wrist strap test is set to 750 kilohms and cannot be modified by the user 5 Insert the ground cord s banana plug in...

Page 4: ...4 Align the backplate s two holes located above and below the tester to the two holes on the pedestal s mounting bracket Secure the backplate using the two included 10 32 pan head screws and nuts 5 Insert one end of the foot plate cord into the foot plate jack located at the bottom of the tester Connect the other end of the foot plate cord into the jack located on the dual foot plate 6 Insert the ...

Page 5: ...Access Control System and Combo Tester X3 Operation 1 A circling light around the test switch indicates when the Combo Tester X3 is on standby and ready to perform a test Access Control System NORMALLY OPEN RELAY DATA INPUT Proximity Reader Combo Tester X3 ACC INPUT Turnstile DATA LINES Figure 8 Wiring the Combo Tester X3 to an Access Control System normally open relay contacts shown above 2 While...

Page 6: ...st strap and footwear to ensure that they are being worn correctly and or need to be replaced NOTE Failures may be caused by dry skin or minimal sweat layer For wrist straps try using an approved dissipative hand lotion such as Menda Reztore ESD Hand Lotion to your wrist prior to use Footwear test results can be improved by taking a short walk to build a sweat layer for better conductivity The Com...

Page 7: ...ar test limit 10 for all other test limits A periodic check using a precision resistance box can be used to verify proper operation The EMIT 50424 Limit Comparator is available for the convenient calibration of the Combo Tester X3 The Limit Comparator allows the customer to perform NIST traceable calibration on the Combo Tester X3 The Limit Comparator can be used on the shop floor within a few min...

Page 8: ...um relative humidity of 80 up to 85 F 30 C decreasing linearly to 50 85 F 30 C Maximum relative humidity of 50 at 1 gigohm setting Dimensions with mounting plate 11 35 x 7 75 x 1 47 288 mm x 197 mm x 37 mm Weight with mounting plate 1 2 lbs 0 5 kg Test Accuracy 20 for 1 gigohm footwear test limit 10 for all other test limits Test Switch Voltage 5 VDC open circuit Wrist Strap and Footwear Test Volt...

Reviews: