- 24 -
y
The primary display will show PASS or FAIL which based on
whether the impedance measured exceeds tolerance range.
The secondary display will show the measurement reading.
y
When sorting mode is active, operate “
SETUP
” key along with
“
Ż
/
Ź
” key, “
Ÿ
/
ź
” key and “
Enter
” key to set the reference
value, range and tolerance.
y
When the setup is finished, press “
ENTER
” key to confirm.
Figure : Setting up the reference value for sorting
Display the measurement reading
Press “
SORTING
” key
The measurement reading is
shown on the secondary display.
Press “
SETUP
” key
“Range” will appear and flash.
Push “
Ż
/
Ź
” key to select the
position of decimal point and units
based on the components you
measure
Press “
ENTER
” key
- 25 -
Last digit will flash
Press “
Ż
/
Ź
” key to select the
desired digit.
Press “
Ÿ
/
ź
” key to adjust value.
Press “
ENTER
” key
“Tol ±” will flash.
Press“
Ż
/
Ź
”key to select Tol.
value : Each press of the key will
select either
±1%,±2%,±5%,±10%,
±20%, -20%+80%, ±0.25% or
±0.5%.
Press “
ENTER
” key
Setup is completed.
Ɣ
If DUT ( device under test ) is
in the specified Tol % , buzzer
beeps once and the primary
display will show “
PASS
”.
y
If DUT ( device under test ) is out
of the specified Tol %, the
primary display will show
“
FAIL
”.