- 20 -
If C < 5pF. The parameter on sub-display is Rp.
*2: When LCR Auto mode is active, the secondary parameter will show
the equivalent resistance in parallel mode (Rp) to replace the D
factor if the C value of DUT is less than 5pF.
*3: LCR AUTO mode only. In Auto-R mode or DCR mode, the
secondary parameter is not available.
4. Inductance / Capacitance / Resistance Measurement
y
Power on the meter, the default testing mode is LCR Auto
mode that can enter to the L/C/R measurement
automatically.
y
Each press of the key will select either Auto-LCR mode,
Auto-L mode, Auto-C mode, Auto-R mode or Auto-DCR
mode.
a. Insert DIP component leads to
the sockets directly. or
b. To measure DIP and SMD
component by
Alligator Test
Lead Case
(TL-21)
If required, Guard line (TL-23) can
provide shield for DUT (device
under test ) preventing from
interference when measuring high
impedance component
Alligator Test Lead
Case (TL-21)
- 21 -
c. To measure SMD component by SMD tweezers (TL-22,
option).
ġ
ɏ
ʳ
To make better precision measurements on extremely high or
low impedance ranges of L,C,R, it is highly recommended
that do OPEN/SHORT calibration before testing.
ġ
Warning : To avoid electric shock, please discharge the capacitor
before testing.
SMD Tweezers
(TL-22)
If required, Guard line (TL-23) can
provide shield for DUT (device
under test ) preventing from
interference when measuring high
impedance component