26
The measurement with method START/STOP is
After the measurement input has recognised a signal ov
values continue to be recorded until the measurement is ended
button. The maximum value is displayed in the field MEASUREMENT.
started when the ON button is clicked.
er the threshold torque, then the
by clicking the OFF
Picture 16: Choosing measurement by START and STOP
UTO is started when ON is clicked. After the
lues
the field MEASUREMENT.
The measurement with method A
measurement
input has recognised a signal over the threshold torque, then the va
continue to be recorded until the current value has fallen short of at least 50% of the
maximum
.
The maximum value is displayed in
Picture 17: Choosing automatic measurement
Note:
When cancelling a measurement procedure,
click with the right mouse button within the
button or option field.
The measurement with method SERIES is st
If you have activated “Show limit input as popup” in the set-up (see
before it ends automatically, double
main window on an area where there is no
arted when ON button is clicked.
Set Up
), a popup-
r the limit values. Enter the current limits either as
ve
window opens to enter the settings fo
upper and lower value or as target value and tolerance in percentage. Click OK to sa
the settings.
IMPORTANT
The first measurement of the series only starts if you have closed the
popup-window.
values cont
The next measurement is st
After the measurement input has recognised a signal over the threshold torque, then the
inue to be recorded until the maximum value has fallen short of at least 50%
of the maximum. The maximum value is displayed in the field MEASUREMENT.
arted again immediately without having to click ON again.