22
Next set the calibration value in the field C-Value. The value can either be entered
sing the arrow keys. An alteration of the calibration value
urement range is displayed again in the lower part of
directly or changed by u
simultaneously causes an alteration of the measurement range.
Now set the correct measurement range. The measurement range display is dependant
n the set unit. The chosen meas
o
the main program (see
Measurement Process
).
11.1.3 External Transducer Settings
When using the external transducer from DEPRAG, choose the option “External-
DEPRAG” on the left-hand side.
Picture 8:
Settings for external transducer
Next set the correct measurement range. The measurement range display is dependant
on the set unit. The chosen measurement range is displayed again in the lower part of
the main program (see
Measurement Process
).
When using a different manufacturer’s external transducer then choose “External-other”.
Picture 9:
Settings for a different manufacturer’s external transducer
Next enter the nominal t
range is displayed again
orque for the corresponding unit. The chosen measurement
in the lower part of the main program (see
Measurement
Process
).
11.1.4 Additional Settings
The amount of measurements in a series, delay and trigger values can be preset on the
right-hand side of the section <Settings>.
The amount of measurements in a series can only be altered if the series measurement
option is chosen (see
Measurement Process
). First choose SERIES. Then the value
can be directly entered or altered using the arrow keys.
The amount of measurements can be preset between 5 and 100.