Alternate method: Place non-bagged instruments
into the steam autoclave. After warm-up is
completed, operate at a sterilizing temperature
and pressure of 273ºF/134ºC to 279ºF/137ºC and
pressure of 31psi/216kPa for 3 to 4 minutes. The
instruments should be used immediately after drying.
Drying:
To dry, use the drying cycle of the autoclave. Set
cycle for 20-30 minutes. Allow instruments to cool
before use.
Maintenance, Inspection and Testing:
• Visually inspect to ensure all contamination is
removed.
• Check for distortion, damage and wear. Discard
damaged, worn or corroded instruments.
Storage:
To maintain sterility, instruments should remain
bagged until ready for use.
In the United States, contact DENTSPLY
Professional Customer Service Technical Support
at 800-989-8826. For areas outside the United
States, contact your local DENTSPLY Professional
Representative.
The Cavitron
®
Steri-Mate™ Light manufactured
by DENTSPLY
®
Professional is warranted against
ENGLISH
MANUFACTURER CONTACT
WARRANTY