4. CARE AND MAINTENANCE

4.1 Hygiene

The SmartLite

®

Max Light is a state-of-the-art electronic device. To prevent unnecessary damage to the unit

and to the power supply, these care and cleaning guidelines must be followed.
• Remove and discard barrier sleeve.

Cross-contamination

• Do not reuse single use products. Dispose of in accordance with local regulations.
• Reprocess reusable products as described below.

• Do not remove curing probe from handpiece until handpiece cleaning and disinfection is completed as

outlined.

4.1.1 Cleaning

• Disconnect the power supply plug from the power outlet and charging base unit prior to

cleaning.

• Clean the unit by gently scrubbing with a disposable towel soaked with hot water and soap or

detergent. Do not allow solution to penetrate the casing.

• Always protect lens of the light probes from abrasive materials. Should the surface become

marred or scratched, light transmission and curing performance will be diminished.

4.1.2 Disinfection

• Probe, handpiece, light shield and/or charging base exposed to spatter or spray of body fluids, or touched

by contaminated hands or oral tissues, should be disinfected with a water-based hospital-level disinfectant
according to national/local regulations. Water-based phenolics are approved disinfectants.

• Do not immerse units in disinfectant. Disinfect the units by spraying disinfectant solution onto a paper towel

and wiping or wiping with impregnated wipes. Do not spray disinfectant directly onto the units or allow
solution to penetrate the casing.

• Use of alcohol, alcohol-containing, iodophor, chlorine or glutaraldehyde based disinfectant solutions will void

the SmartLite

®

Max Light warranty.

• DO NOT immerse unit in water or disinfectant. Wipe – do not spray – solution (with curing probe in place)

onto plastic parts. Prevent liquids from entering openings on unit.

• DO NOT autoclave or otherwise sterilize these components.
• Reconnect the power supply plug to the power outlet and the charging base unit when cleaning and

disinfection process is completed.

4.2 Head replacement

The head portion of the SmartLite

®

Max Light may be removed for replacement or cleaning. To do this, simply

grasp the head firmly and pull it away from the lower handle portion. Attach the replacement head by inserting
the connector into the top of the handle. Make sure you push the head all the way against the handle to ensure
a secure connection. Make sure the power supply is disconnected from the unit before replacing the head.

5. TROUBLESHOOTING

PROBLEM: There is no display.

CAUSE: The unit is in hibernation mode.
SOLUTION: Press the on/off button to wake the unit.

PROBLEM: The unit does not appear to have any power.

CAUSE: The battery is drained.
SOLUTION: Attach the power supply to the SmartLite

®

Max Light handle, or place the SmartLite

®

Max Light in

the charging cradle, making sure the power supply is attached to the cradle.

CAUSE: The power supply is not attached to the SmartLite

®

Max Light or the electrical outlet.

SOLUTION: Make sure the power supply is securely connected to electrical supply outlet and to the

SmartLite

®

Max Light handle or cradle.

CAUSE: The SmartLite

®

Max Light is in sleep mode.

SOLUTION: Press any button once to bring the unit out of sleep mode. The display should illuminate when this

is done.

Summary of Contents for SmartLite Max

Page 1: ...DIRECTIONS FOR USE...

Page 2: ...TABLE OF CONTENTS FOR MULTI LANGUAGE...

Page 3: ...ower supply can be attached if needed ensuring consistent uninterrupted use the entire day The SmartLite Max Light also features a fully rotating head with a complete 360 range of movement allowing yo...

Page 4: ...ut Curing Light 12VDC 2 0A Power Input Power Supply 90 240VAC 50 60 Hz Pulse Rate 10 Hz Pulse Duty Cycle 50 Operating Temperature 15 30 C 59 86 F 0 95 RH 1000 8000 meters elevation 2 GENERAL SAFETY NO...

Page 5: ...emitted from the curing probe As with any bright light source staring directly at the source may cause temporary effects such as after image See Warnings 3 It is recommended that all curing light use...

Page 6: ...Light check the packaging and parts for any possible damage that may have occurred in transit If damage is apparent please contact your authorized DENTSPLY distributor from which the light was purchas...

Page 7: ...utput is pulsed 10 times a second giving a net effect of half the power output as the continuous mode Ramp There is a steady increase of the output from 0 to 1200mW cm2 within 4 seconds and then remai...

Page 8: ...buttons to select the desired settings Since the probe can be rotated 360 it could be used both as a wand style with thumb operation or gun style with index finger operation Always use the settings re...

Page 9: ...lcohol alcohol containing iodophor chlorine or glutaraldehyde based disinfectant solutions will void the SmartLite Max Light warranty DO NOT immerse unit in water or disinfectant Wipe do not spray sol...

Page 10: ...me PROBLEM I press Start but no light comes out CAUSE Head is not attached properly SOLUTION Make sure the head is pushed completely into the handle such that the head connector is not visible PROBLEM...

Page 11: ...oduct DAMAGES ARE LIMITED STRICTLY TO REPLACEMENT OF THE PRODUCT DENTSPLY EXPRESSLY DISCLAIMS LIABILITY FOR INCIDENTAL AND CONSEQUENTIAL DAMAGES RESULTING FROM ITS USE Claims covered by this warranty...

Page 12: ...ENTSPLY Sankin K K 14 9 Yushima 3 chrome Bunkyo ku Tokyo 113 0034 Japan Tel 81 03 5114 1001 Hong Kong Division of DENTSPLY International 25 F Pacific Link Tower Southmark Tower A 11 Yip Hing Street Wo...

Reviews: