PROBLEM: The head or handle is getting hot.

CAUSE: Unit has been used for too many consecutive cycles.
SOLUTION: Place the SmartLite

®

Max Light in the charger, or set the unit on a counter, and allow it to cool

down for at least 6 minutes.

PROBLEM: Power output indicated on the cradle does not match the results from another power meter.

CAUSE: The cradle power meter is not a quantitative measuring device.
SOLUTION: The power meter incorporated into the charging cradle is meant as a simple diagnostic tool only,

and is calibrated to the output performance of the SmartLite

®

Max Light. Other curing lights may

not give a similar result on the SmartLite

®

Max Light power meter as the SmartLite

®

Max Light

does. Due to the oval shape and dispersion angle of the SmartLite

®

Max Light, the power

measured on another power meter may not be accurate.

PROBLEM: The power meter doesn’t show a green bar when the output is measured.

CAUSE: SmartLite

®

Max Light head is not placed in the sensor correctly.

SOLUTION: Make sure the SmartLite

®

Max Light head is resting securely in the sensor, with the head resting

against the sensor face.

CAUSE: SmartLite

®

Max Light is in Pulse mode.

SOLUTION: Do not measure the power output in Pulse mode. Select another mode to measure the power output.

PROBLEM: I can’t select the time in Boost mode.

CAUSE: Timer function is not available in Boost mode.
SOLUTION: If more time is desired, either use the Boost mode repeatedly at the 5 second curing cycle, or

select another function and choose the desired curing time.

PROBLEM: I press Start, but no light comes out.

CAUSE: Head is not attached properly.
SOLUTION: Make sure the head is pushed completely into the handle, such that the head connector is not

visible.

PROBLEM: I’ve been using the unit for several cycles and it suddenly stopped working.

CAUSE: Temperature overload – sensor has activated.
SOLUTION: The SmartLite

®

Max Light is equipped with a temperature overload sensor to protect the LEDs

from excessive heat. If this occurs, place the unit in the cradle or on the counter and wait at least
6 minutes for the unit to cool.

CAUSE: Battery has been drained.
SOLUTION: The battery has been drained of its stored energy. Place the unit in the charging cradle, or attach
the power supply to the handle.

PROBLEM: The red battery low indicator light is on and I plugged in the power cord, but the light still

won’t function.

CAUSE: The light has to reset when the battery low indicator light is on continuously and the power cord is

plugged in for corded use.

SOLUTION: plug in the power cord and let the light reset for several minutes before use.

If the above steps do not solve the problem, contact the dealer or point of purchase. DO NOT USE THE
SmartLite

®

Max Light AND DO NOT ATTEMPT TO SERVICE THE UNIT. Serious injury may result.

SmartLite

®

Max L.E.D. Curing Light complies with the following Directives/Standards:

• Medical electrical equipment Part 1: General requirements for safety (IEC 60601-1)
• Standard for Safety for Medical Electrical Equipment, Part 1: General Requirements for Safety (UL 60601-1

and CAN/CSA C22.2 No. 601.1-M90)

• Medical Electrical Equipment - Part 1-4: General Requirements for Safety - Collateral Standard:

Programmable Electrical Medical Systems Consolidated (IEC 60601-1-4)

8

Summary of Contents for SmartLite Max

Page 1: ...DIRECTIONS FOR USE...

Page 2: ...TABLE OF CONTENTS FOR MULTI LANGUAGE...

Page 3: ...ower supply can be attached if needed ensuring consistent uninterrupted use the entire day The SmartLite Max Light also features a fully rotating head with a complete 360 range of movement allowing yo...

Page 4: ...ut Curing Light 12VDC 2 0A Power Input Power Supply 90 240VAC 50 60 Hz Pulse Rate 10 Hz Pulse Duty Cycle 50 Operating Temperature 15 30 C 59 86 F 0 95 RH 1000 8000 meters elevation 2 GENERAL SAFETY NO...

Page 5: ...emitted from the curing probe As with any bright light source staring directly at the source may cause temporary effects such as after image See Warnings 3 It is recommended that all curing light use...

Page 6: ...Light check the packaging and parts for any possible damage that may have occurred in transit If damage is apparent please contact your authorized DENTSPLY distributor from which the light was purchas...

Page 7: ...utput is pulsed 10 times a second giving a net effect of half the power output as the continuous mode Ramp There is a steady increase of the output from 0 to 1200mW cm2 within 4 seconds and then remai...

Page 8: ...buttons to select the desired settings Since the probe can be rotated 360 it could be used both as a wand style with thumb operation or gun style with index finger operation Always use the settings re...

Page 9: ...lcohol alcohol containing iodophor chlorine or glutaraldehyde based disinfectant solutions will void the SmartLite Max Light warranty DO NOT immerse unit in water or disinfectant Wipe do not spray sol...

Page 10: ...me PROBLEM I press Start but no light comes out CAUSE Head is not attached properly SOLUTION Make sure the head is pushed completely into the handle such that the head connector is not visible PROBLEM...

Page 11: ...oduct DAMAGES ARE LIMITED STRICTLY TO REPLACEMENT OF THE PRODUCT DENTSPLY EXPRESSLY DISCLAIMS LIABILITY FOR INCIDENTAL AND CONSEQUENTIAL DAMAGES RESULTING FROM ITS USE Claims covered by this warranty...

Page 12: ...ENTSPLY Sankin K K 14 9 Yushima 3 chrome Bunkyo ku Tokyo 113 0034 Japan Tel 81 03 5114 1001 Hong Kong Division of DENTSPLY International 25 F Pacific Link Tower Southmark Tower A 11 Yip Hing Street Wo...

Reviews: