29
Maintenance of Unit 8
8.1 Draining the filter
Check the filter bowl from the filter inspection window, and
if there is an accumulation of water, insert fingers from the
bottom, and turn the drain cock in the direction of the arrow
to drain water from the filter (Fig. 23). After draining, securely
tighten the drain cock by turning it in the reverse direction.
8.2 Changing O-ring (Use of Water Supply Bottle)
Remove the two O-rings at the bottle joint with a pointed
tool, and mount a new nose into the grooves (Fig. 24).
Optional O-ring: Order No.Y900-327
8.3 Changing Water Filter (Use of City Water)
1.
Close the water valve of to connect the Cercon
®
mill.
Mount two (5 x 8) spanner wrenches as shown in Fig. 25,
and turn in the direction shown. When the water tube be-
comes twisted as a result of this, relieve the twisting by
turning it, as its end on the control unit side is free to turn.
2.
When the water filter case is separated, the water filter
can be removed as shown in Fig. 26. Replace with a new
one and reassemble the filter in the reverse order (Fig. 26).
Water Filter: Order No. U387-042
Fig. 23
• Assemble the water filter in the correct
direction as shown in Fig. 26.
Fig. 24
Fig. 25
Fig. 26
Water Filter
O-ring
Control Unit Side
Water Filter Case
O-ring
Summary of Contents for DeguDent Cercon mill
Page 2: ......
Page 17: ...17 ...
Page 31: ...31 ...
Page 45: ...45 ...
Page 59: ...59 ...
Page 73: ...73 ...
Page 87: ...87 ...
Page 101: ...101 ...
Page 115: ...115 ...
Page 129: ...129 ...
Page 143: ...143 ...
Page 157: ...157 ...
Page 171: ...171 ...
Page 185: ...185 ...
Page 199: ...199 ...
Page 213: ...213 ...
Page 227: ......
Page 241: ...241 ...
Page 242: ......
Page 243: ......