2 Specifications
Speed:
Proper Air Supply Pressure:
320.000 min
-1
0,25~0,35 MPa
Handpiece Dimensions:
Handpiece Weight:
Ø 16,6 mm x H 136,6 mm, 0.65 in/5.37 in
104 g (including coupling joint), 3.67 oz
Control Unit Dimensions:
Control Unit Weight:
W 214 mm x D 110 mm x H 175 mm
1,3 kg, 2.86 Ib
W 8.42 in/D 4.33 in/H 6.89 in
3 Installation Procedures of Each Component
3.1 Installation of Plumbing Hose
Push the plumbing hose into the plumbing connector
located at the regulator on the left-hand side of the control
unit until it is securely set to make connection as shown
in Fig. 1. Check if the hose is securely installed by slightly
tugging it after connection. Connect the other end of the
plumbing hose to the air line. At this time, use the attached
connector if necessary.
• Push the plumbing hose into the connec-
tor until it is securely set. Otherwise, air
may leak.
• Pushing the white ring, on the plumbing
connector, gently remove the tube.
3.2 Installation of Foot Pedal
Loosen and remove the hose stoppers (two) from the foot
pedal connectors on the back of the control unit, and in-
sert them into the foot pedal hoses. At this time, insert the
hose stoppers so that their screws come outside (control
unit side). Insert the „1“-marked end of each hose accord-
ing to the instructions on the label. After insertion of both
hoses, securely tighten the hose stoppers (Fig. 2).
22
Fig. 1
Fig. 2
Foot Pedal Connector
Hose Stopper
Plumbing Hose
Plumbing Connector
White Ring
Summary of Contents for DeguDent Cercon mill
Page 2: ......
Page 17: ...17 ...
Page 31: ...31 ...
Page 45: ...45 ...
Page 59: ...59 ...
Page 73: ...73 ...
Page 87: ...87 ...
Page 101: ...101 ...
Page 115: ...115 ...
Page 129: ...129 ...
Page 143: ...143 ...
Page 157: ...157 ...
Page 171: ...171 ...
Page 185: ...185 ...
Page 199: ...199 ...
Page 213: ...213 ...
Page 227: ......
Page 241: ...241 ...
Page 242: ......
Page 243: ......