21
Maintenance
For optimal performance and device life, maintenance of the Calamus
®
Dual
should be performed at least every 12 months.
FLOW Cartridge Chamber
Over time gutta-percha can build up or become trapped in the cartridge
chamber of the FLOW Handpiece. This is normal.
To Remove:
1.
Turn on the unit and press the
button on the console to fully
retract the plunger.
2.
Allow the heater section to reach operation temperature (180°C).
3.
Turn off unit.
4.
Insert the included cleaning brush into the heating chamber. Rotate
the brush several times to remove the gutta-percha from the
chamber.
FLOW Handpiece Plunger
1.
Without a cartridge inserted, press the 360° activation cuff until the
gutta-percha indicator reaches the top of the chamber.
2.
Then, push the
button and allow the indicator to completely
retract.