7
• Visually inspect all o-rings for wear. Replace worn o-rings
(Refer to O-Ring Replacement Kit Instructions). Lubricate
o-rings with water before use.
• Check for distortion, damage and wear (Refer to Efficiency
Wear Indicator Instructions). Discard damaged, worn or
corroded instruments.
Storage:
To maintain sterility, instruments should remain bagged until
ready for use.
1) For cards with notches, place the insert tip flat against the
applicable Insert Tip diagram with the engraved line aligned
with the edge of the card. For cards without notches, match
the insert to the appropriate graphic. Hold the top of the
grip flush against the edge of the wear guide. Rotate the
insert so the tip is flat against the card.
2) Compare length to the red and blue lines.
3) Evaluate tip shape. If shape does not match profile, discard
the insert.
To re-order Inserts, call your local DENTSPLY Professional
Distributor or call Customer Service 800-989-8826
To order efficiency indicator cards for Cavitron
®
inserts: Call
DENTSPLY Professional Customer Service at 1-800-989-8826.
Cavitron
®
ultrasonic inserts manufactured by DENTSPLY
Professional are warranted against defects arising from faulty
ENGLISH
HOW TO USE THE EFFICIENCY WEAR INDICATOR
Tip touches
blue line
Reorder
Tip touches
red line
Discard
Tip extends beyond
blue line
Optimal Efficiency
Red Blue
WARRANTY
Summary of Contents for Cavitron
Page 43: ...43 ITALIANO...
Page 44: ...44 Cavitron Cavitron Cavitron Centers for Disease Control and Prevention CDC...
Page 45: ...45 Cavitron 35 Cavitron Cavitron Cavitron Cavitron...
Page 47: ...47 1 2 3 4 5 6 7 8 9 Cavitron...
Page 48: ...48 15...
Page 49: ...49 135 C 275 F 3 5 132 C 270 F 3 20 30...
Page 50: ...50 1 c 2 3...
Page 52: ...52 4 DENTSPLY DENTSPLY Cavitron DENTSPLY Professional 1 800 989 8826...
Page 53: ...53...
Page 54: ...54...
Page 55: ...55...