Dentsply Sirona
Operating Instructions SiroCam UAF Plus
2 Safety
1.5 Requirements
66 90 643 D3762
D3762.201.01.01.02 08.2018
9
2
Safety
Check the Camera before Using It
There are no customer-serviceable components in the camera.
However, before each usage, check the outer surface of the camera for
any signs of physical damage or defect. The surface of the camera
should have a smooth finish, with no evidence of chipping or damage to
either the handpiece housing or the lens section. If detected, contact
your local distributor of Denstply Sirona products for further instructions.
Always Use Sheaths
Use protective camera sheaths every time the camera is used. Never
use the intraoral camera without a protective sheath. Never use a
damaged sheath. Always dispose of the sheath after every patient.
Protective sheaths are single-use items and must not be reused under
any circumstance. Reuse of single-use devices/instruments may cause
them to become contaminated, compromise their intended function, and
result in patient and user infection, injury and/or illness.
RF Interference Considerations
Although the camera is designed to provide a reasonable degree of
protection from electromagnetic interference, according to IEC
International regulations, it must be installed at an adequate distance
from electricity transformer rooms, static continuity units, two-way
amateur radios and cellular phones. To ensure proper operation, the
latter (meaning, electricity transformer rooms, static continuity units,
two-way amateur radios and cellular phones) can be used only at a
minimum distance of 1.5m from any part of the camera.
Any instrumentation or equipment for professional use located near the
camera must conform to Electromagnetic Compatibility regulations.
Nonconforming equipment, with known poor immunity to
electromagnetic fields, may not operate properly unless they are
installed at a distance of at least 3m and supplied by a dedicated
electric line.
Operate the Camera as Directed
Always use the camera in accordance with the directions and
recommendations contained in this guide and other accompanying or
referenced documentation. Do not attempt to modify the camera or use
it in system configurations not specified in this or other product
documents.
Only Dentists or Authorized Designees Are Permitted to Operate the
System
To ensure the correct use of the camera in a clinical environment, for
purposes that correspond to its intended design and application, only
dentists, or their designees, are authorized to operate the device.
Ensure Proper Camera and PC Workstation Installation and
Operation
The camera has been determined to be in accordance with international
safety standards and is deemed suitable for use within the patient area,