2 Safety
2.1 Symbols
Dentsply Sirona
Operating Instructions SiroCam UAF Plus
10
66 90 643 D3762
D3762.201.01.01.02 08.2018
which extends from the patient for a distance of 1.5m. To comply with
these standards, do not operate non-medical equipment (such as a PC
workstation) inside the patient area. Outside the patient area, the
presence of approved non-medical grade equipment and Listed /
Approved / IEC 60950-1 certified Information Technology Equipment
(ITE) computer equipment is acceptable.
The host computer (PC workstation) should be CE-approved and
conform with the Low Voltage [73/23/EC] and EMC Directive [89/336/
ERC]. Also, to help ensure optimal performance, ensure that all
software programs residing on the workstation are virus-free and have
been adequately tested so they will not impact imaging applications
after installation.
2.1
Symbols
Sirona Dental products display markings to indicate compliance with
regulatory requirements and applicable technical standards.
Indicates Type BF equipment in accordance with applicable medical
device safety standards (IEC/EN/UL 60601-1)
Indicates an attention to customers to consult accompanying
documents for information needed for proper use of device
Manufacturer’s address
Indicates conformance to European Union Medical Devices Directive
(MDD) 93/42/EEC
Signifies certification to the product safety requirements of the USA and
Canada
These symbols, with the serial number of the camera, can be found on
the label attached to the the camera cable.