Automatic Shutdown
ProPex
®
II
automatically shuts down after 5 minutes without use. It is advisable,
however, to manually switch off equipment after measurement by simply pressing the
key (On/Off).
I.
Maintenance of your ProPex
®
II
The device does not contain user serviceable parts. The service and repair should be
provided by factory trained service personnel only.
After each use, all the objects that were in contact with infectious agents should be
cleaned using towels impregnated with a disinfecting and detergent solution (a
bactericidal, fungicidal and aldehyde free solution). Use of chemical agents may cause
damage to the equipment.
Reprocessing procedure :
-
the lip clip, the hook and the fork must be sterilized between treatments by
autoclaving at 134°C
See after for detailed procedure
The measuring cable can not be autoclaved.
J.
Warranty
ProPex
®
II
is warranted for 24 months from the date of purchase. The accessories
(cables, battery etc.) are warranted for 6 months from the date of purchase.
The warranty is valid for normal usage conditions. Any modification or accidental damage
will render the warranty void.