![Dentsply Sirona Midwest Tradition Pro TBF Operating Instructions Manual Download Page 25](http://html1.mh-extra.com/html/dentsply-sirona/midwest-tradition-pro-tbf/midwest-tradition-pro-tbf_operating-instructions-manual_2483680025.webp)
Operating instructions
3 Technical description
66 27 587 D3673.201.02.04.02
23
English
Midwest Tradition TB /
Pro TBF / TL / Pro TLF
Spray air flow in Nl/min
> 2.5
Spray water pressure in bar (psi)
2.0 ± 0.2 (29)
Water content in spray in ml/min
> 50
The indicated pressures can be checked by a service engineer with
a star tester that is placed between the supply hose and the high-
speed handpiece.