8 Maintenance
Operating Instructions Midwest Stylus
®
Plus / Midwest
®
Phoenix ™ /Pro
64
6607290 D3673
.
201.01.03.02
8.3
Servicing Midwest
®
quick coupling
8.3.1
Replacing the halogen lamp of the 6-pin coupling
CAUTION! The lamp can be hot. Risk of burns! Allow the lamp to cool
down.
1. Remove the high-speed handpiece.
2. Insert blunt pick into recessed area on lamp base and slide lamp out
of socket.
3. Remove the defective lamp (A).
IMPORTANT: Do not touch replacement lamp with bare fingers.
Oils from the skin decrease lamp life. Use gloves when handling
lamp.
4. Insert carefully the new lamp with contact pin end first into lamp
socket.
Be sure to orient recessed area on lamp base to face out slot on
A
6607290.book Seite 64 Mittwoch, 13. September 2017 9:53 09