5 Operation
Operating Instructions Midwest Stylus
®
Plus / Midwest
®
Phoenix ™ /Pro
40
6607290 D3673
.
201.01.03.02
5.4
Adjusting the cooling spray
> 50 ml/min
➢ Adjust the flow rate of the cooling water using the control ring (A) (>
50 ml/min).
Tip: You can measure the amount of cooling water with a measuring
cup and watch.
Water flow rate
Quick coupling F
Quick coupling F (Dentsply Sirona): The maximum water flow is set
when the control ring clicks into place.
KaVo quick coupling
Multiflex LUX quick coupling (KaVo): The maximum water flow is set
when the two markings face each other.
NSK MachLite quick coupling
MachLite quick coupling (NSK): The maximum water flow is set when
the control ring clicks audibly into place.
A