7 Conditioning
Operating Instructions Midwest Stylus
®
Plus / Midwest
®
Phoenix ™ /Pro
46
6607290 D3673
.
201.01.03.02
7.3
Manual cleaning and disinfection
IMPORTANT: Manual conditioning is possible in exceptional cases if
the national/regional regulations are followed. The national/regional
regulations are to be checked before.
NOTICE! Condition immediately, or at the latest, one hour after
treatment.
NOTICE! Never clean in an ultrasound bath!
NOTICE! Never immerse in disinfectant solution!
NOTICE! Only use Midwest
®
Plus Aerosol Spray.
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
✔ Wear appropriate protective clothing.
✔ All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.
6607290.book Seite 46 Mittwoch, 13. September 2017 9:53 09