Dentsply Sirona
Instructions for CEREC Bluecam and CEREC Omnicam
2 CEREC Omnicam
2.8 High-level disinfection of the mirror sleeve with the HLD set 63 46 907
62 82 839 D3492
D3492.201.02.06.02 10.2018
11
17. Dry the entire outer surface of the mirror sleeve as well as the inner
section using a soft lint-free cloth. Ensure that no fluid is able to
penetrate the inside of the mirror sleeve.
18. Store the mirror sleeve in such a way that it is protected from
contamination until the next use.
19. Remove the white protective cap from the camera head before use.
20. Carefully re-attach the mirror sleeve and allow it to lock in place
(see “Refitting the mirror sleeve [→ 5]”).
2.8
High-level disinfection of the mirror sleeve with
the HLD set 63 46 907
NOTE
The hot air sterilization and high-level disinfection must not be
combined.
The complete process for high-level disinfection (HLD) is as follows –
assuming the CIDEX
®
OPA disinfectant is available in your country via
your dealer:
1.
Carry out a preliminary cleaning process (see "Pre-cleaning of the
mirror sleeve [→ 6]").
2.
Always use personal protective equipment when implementing the
high-level disinfection if not before.
3.
Remove the mirror sleeve from the CEREC Omnicam (see
"Removing the mirror sleeve [→ 5]").
4.
Place the white protective cap on the camera head and place the
camera in the camera storage unit.
5.
Use the following disinfectant for the high-level disinfection: CIDEX
®
OPA.
CAUTION
Observe the safety instructions from the disinfectant manufacturer.