Dentsply Sirona
Instructions for CEREC Bluecam and CEREC Omnicam
2 CEREC Omnicam
2.7 High-level disinfection of the mirror sleeve with the HLD set 66 05 120
62 82 839 D3492
D3492.201.02.06.02 10.2018
7
2.7
High-level disinfection of the mirror sleeve with
the HLD set 66 05 120
NOTE
The hot air sterilization and high-level disinfection must not be
combined.
The complete process for high-level disinfection (HLD) is as follows –
assuming the CIDEX
®
OPA disinfectant is available in your country via
your dealer:
A
HLD tank
C
Metal bracket
B
Cover
D
Plugs
1.
Carry out a preliminary cleaning process (see "Pre-cleaning of the
mirror sleeve [→ 6]").
2.
Always use personal protective equipment when implementing the
high-level disinfection if not before.
3.
Remove the mirror sleeve from the CEREC Omnicam (see
“Removing the mirror sleeve [→ 5]”).
4.
Place the white protective cap on the camera head and place the
camera in the camera storage unit.
5.
Use the following disinfectant for the high-level disinfection: CIDEX
®
OPA.