8 Maintenance
8.3 Color calibration
Dentsply Sirona
Operating Instructions CEREC AC Connect
68
64 32 087 D3492
D3492.201.07.24.02 07.2018
Color-calibrating the camera
1.
Make sure that the CEREC Omnicam is clean, disinfected and dry.
2.
Remove the color-calibration set from the packaging.
3.
Use the CEREC Omnicam to scan the QR code on the underside of
your color-calibration set. In order to do this, you must hold the
CEREC Omnicam still in front of the QR code so that it is
completely visible in the picture. If the QR code appears to be shiny,
hold the camera at more of an oblique angle in order to avoid any
glaring light and to make it easier to scan the codes. If the QR code
is recognized, the next
"Please mount color calibration set" step
appears.
This step of the QR code scan is skipped during the subsequent
color calibration and the serial number of the color-calibration set is
thus displayed. If this does not match the serial number printed on
your color-calibration set, click on the
"Rescan QR Code" button
and scan the new QR code.
4.
Mount the color-calibration set on the tip of the camera until it locks
into place.
5.
Click on the
"Ok" button.
Ä
The measuring process starts. Do not move the CEREC
Omnicam or the color-calibration set during this time.
Ä
The software provides status updates on the calibration and
informs you once the procedure is complete.
Ending the color calibration
ü
The software indicates that the color calibration was completed
successfully.
1.
Click on the
"Ok" button.
Ä
The CEREC Omnicam is now color calibrated.
2.
Remove the color-calibration set from the camera and place it back
in the packaging.
Error message during color calibration
The software indicates if an error occurs during color calibration. If the
color calibration contained an error, ensure the following:
● The color-calibration set is free of dust
● The color-calibration set was mounted correctly
● The CEREC Omnicam exit window is clean
➢ Then restart the color calibration.
Do not continue using a damaged color-calibration set; instead, contact
your distributor to purchase a new one.
Summary of Contents for CEREC AC
Page 91: ......