8 Maintenance
8.1 Care, cleaning, disinfection, and sterilization
Dentsply Sirona
Operating Instructions CEREC AC Connect
60
64 32 087 D3492
D3492.201.07.24.02 07.2018
11. Remove the cover and take the bracket with the mirror sleeve out of
the container.
12. Carefully rinse the mirror sleeve with water from the mains while
holding it with the bracket.
13. Remove the bracket from the mirror sleeve carefully and slowly by
pushing the bracket downwards. Ensure that the mirror sleeve is
pointing upwards so that no fluids are able to penetrate into the
interior of the mirror sleeve.
14. Remove the plug from the bracket.
15. Dry the plug.
16. Dispose of the fluid and store the bracket in the empty container if
the HLD set is not used for more than one week.