6 Maintenance and repair
6.3 Using the cleaning hose and the wet cleaning process
Dentsply Sirona
Operating Instructions for suction devices
22
66 25 540 D3608
D3608.201.02.05.02 10.2020
CEREC Primemill: Open the door of the production chamber and
activate the cleaning position in the
"Routine Actions" area of the
touch interface in order to bring the motors together so that the
production chamber is easier to access and clean. This function can
be started via the
"Routine Actions" area (symbol on left) by
selecting the
"Cleaning Position" item.
2.
Rotate the black adapter of the manual cleaning hose set from
suction of the chamber to suction through the manual cleaning
hose.
3.
Activate the manual operation for the suction by setting the switch
at the top of the device from "
Auto
" to "
On
".
4.
Suck up the dust from the chamber as required.
5.
Sucking up as much dust as possible from the workpiece axis is
recommended. Also remove the dust from the bottom side of the
block axis as well as on the bearing units and scrapers of the right
and left machining axes (see arrows).
6.
Close the door.
Ä
The motor mounts and the tool axis move back to initial
position.
7.
CEREC MC XL / MC X / MC: Once the chamber has been cleaned
using the suction device, you should complete the 30-second wet
cleaning process programmed in the software 2 to 3 times as
required in order to remove any accumulated material.
- In the system menu of the software, navigate to
"Configuration"
and click on the button
"Devices".
- Select the corresponding unit.
- Click on the option
"Service" and then on "Start a cleaning
process.". Alternatively, you can also wet grind a normal restoration
with the same effect.
CEREC Primemill: After vacuuming the chamber, you should
perform the wet cleaning of the machine. This function can be
started via the
"Routine Actions" area (symbol on left) by selecting
the
"Cleaning Program" item. There are two cleaning options: 2
minutes and 15 minutes. Each cleaning option can be stopped at
any time when the desired results have been achieved.
8.
You can remove any zirconium oxide deposits in the lower part of
the workpiece axis with a nylon brush between the wet cleaning
processes.
Summary of Contents for 6385129
Page 2: ......
Page 29: ......
Page 87: ......
Page 89: ...10 2020 inLab MC X5 CEREC Primemill CEREC MC MC X MC XL Premium Package...
Page 96: ...1 1 6 Dentsply Sirona 8 66 25 540 D3608 D3608 201 02 05 12 10 2020 1 6 20XX 28...
Page 111: ...Dentsply Sirona 6 6 2 66 25 540 D3608 D3608 201 02 05 12 10 2020 23 5 6 HEPA 7 8 9 6 2 800...
Page 114: ...7 7 1 Dentsply Sirona 26 66 25 540 D3608 D3608 201 02 05 12 10 2020 7 7 1 HEPA HEPA...
Page 115: ...Dentsply Sirona 7 7 2 66 25 540 D3608 D3608 201 02 05 12 10 2020 27 7 2 14 HEPA HEPA HEPA HEPA...
Page 119: ...10 2020 inLab MC X5 CEREC Primemill CEREC MC MC X MC XL Premium Package...
Page 125: ...Dentsply Sirona 1 1 6 66 25 540 D3608 D3608 201 02 05 27 10 2020 7 1 6 20XX 26...
Page 126: ...1 1 6 Dentsply Sirona 8 66 25 540 D3608 D3608 201 02 05 27 10 2020...
Page 142: ...7 7 1 Dentsply Sirona 24 66 25 540 D3608 D3608 201 02 05 27 10 2020 7 7 1 On min max HEPA HEPA...
Page 147: ......