Cleaning, Disinfection and Sterilization
60/74
F1902142.EN / 05 / 2015 / updated 02/2016
Contra-angle trial run after lubrication:
N°
Action
A
Attach the contra-angle to the motor handpiece.
B
Hold the motor handpiece upside down.
C
Run the motor handpiece for 30 seconds to remove any excess oil.
D
If dirt emerges, repeat the entire cleaning and disinfection process and start
the lubrication process again.
E
Wipe the contra-angle with a paper tissue or a soft cloth.
Always calibrate the motor handpiece when the contra-angle has
been lubricated or has been exchanged after sterilization.
See
for the calibration function from the
DENTSPLY ENDO IQ™
application.
Calibrate the motor handpiece without any files attached.
If you only use the motor handpiece in reciprocating mode (without
the
DENTSPLY ENDO IQ™
application), no calibration is required.