Warnings
10/74
F1902142.EN / 05 / 2015 / updated 02/2016
•
The iPad
®
protective case and the motor handpiece cover contain
permanent magnets. Magnets could affect the normal functioning of
pacemakers, implanted heart defibrillators and hearing aid devices.
Do not place the accessories near these devices or any other object
sensitive to static magnetic fields.
•
Do not charge, operate or store at high temperatures. Comply with
the specified operating and storage conditions.
During treatment
•
Gloves and a rubber dam are absolutely essential during treatment.
•
If any irregularities with the device should occur during treatment,
switch it off. Contact Dentsply Maillefer or your Dentsply Maillefer
authorized service partner.
Battery
•
Always use the original charger with the charging cable to charge
the device. Use of non-original chargers jeopardizes the safety of
the patient and user.
•
If liquid should emerge from the motor handpiece, the batteries may
be leaking. In this event, stop using the device immediately and
contact Dentsply Maillefer or your Dentsply Maillefer authorized
service partner
.
•
Never open the device yourself. Opening the device voids the
warranty. Contact Dentsply Maillefer or your Dentsply Maillefer
authorized service partner
.
•
A device with a faulty battery must not be sent by air freight
.
•
Always check to ensure that both the motor handpiece and the
iPad
®
have a sufficient battery charge before starting treatment.
Contra-angle
•
Never press the contra-angle pushbutton when the motor
handpiece is running or if it is coming to a stop. This will lead to
detachment of the instrument or cause the pushbutton to overheat.
•
Never remove the contra-angle from the motor handpiece during
operation.
•
Only use undamaged root canal instruments. Please refer to the
information provided by the manufacturer.