21 / 112
Operation
Operating mode
Warning
9. Sterilization
- Steam sterilization at:
134°C (237°F) during
3 min
- The accessories (lip clip, hook
and fork) must be sterilized
according to the packaging
labeling.
- Use fractionated vacuum or
gravity (less preferred) autoclaves
(according to EN 13060, EN 285).
- Use validated sterilization
procedure according to ISO
17665-1
- Respect maintenance procedure
of the autoclave device given by
the manufacturer.
- Use only the listed sterilization
procedures.
10. Storage
- Keep devices in
sterilization packaging in
a dry and clean
environment
- Sterility cannot be guaranteed if
packaging is open, damaged or
wet (check the packaging before
using the instruments).