counter-clockwise and unscrew Lens Module. Replace damaged lens with a new Lens Module by reversing the
steps. Do not overtight the thread and O-rings underneath.
Barrier Sleeves:
Slide sleeve into Handpiece. Twist body relative to sleeve for tight fit.
Light Shield:
Slide in Light Shield from the LED Light Head to the Handpiece into proper position for best access.
Tip Size:
Tip and beam size can be adjusted by selecting the proper size Curing Caps. For
small cavity treatment, snap on 9mm Curing Cap and anchor light on occlusal or bonding
surface. For temporary bonding, snap on 3mm Curing Cap (Tacking Tip) at the end of the
Lens Module and apply pressure at the center of crown/veneer for best contact. Tack for 1
second then remove excess adhesives before final bonding. For veneer bonding and molar
cavity treatment, snap on 13mm Curing Cap and place light in direct contact.
Cleaning and Disinfecting
Clean and disinfect Handpiece and Charging Stand with a soft cloth or tissue dampened with cleaning solution
including: Cavicide, Lysol, Isopropyl or Ethyl alcohol and equivalent. Do not spray, soak, or autoclave.
Use Q-tips to clean lens and electrical contacts by slightly dampening the tip (do not soak) with Isopropyl alcohol.
Safeguard Handpiece opening area, Metal Keypad, Charging Plug, and Lens Module against liquid and sharp
objects as this may cause electrical short-circuit, permanent optical damage or malfunction.
Optional Add-ons
FUSION-DOE is the most advanced and versatile LED light system allowing users to perform beyond normal and
composite curing and exams. You can implement additional advanced curing and therapeutic procedures by twisting
and pulling out its modular light head and switch it for custom illumination. Additional optional light heads are:
Plus Head
–
providing larger (12mm) curing tip with broader wavelength coverage from
385 to 490nm. It allows curing of larger areas than the standard 10mm tip and cures wider
range of photoinitiators.
Twinhead
–
Trans-enamel cure: Rotate each head to approximately face each other and
place light heads from both the buccal and the lingual at the same time. Set timer to 3
seconds in Plasma mode (cure twice) or 10 seconds in Pulse mode and cure. Help
eliminating shrinkage (e.g. bulk cure) or indirect brackets bonding. Rotate each head to point it at a different spot
from the other head also allows curing of multiple teeth and brackets. Popping out one head the same way as the
Standard LED Light Head enables the remaining head to cure in full power of 4000mW/cm2 in Plasma Mode.
fPad Light Head
- providing red light (620
–
650 nm) for 1) photodynamic pain therapy for local pain relief for TMJ
patient or after oral surgery; 2) photo-activated disinfection of gum diseases.
Product Specification
Type of information
Specifications
Wavelength
LED Light Head
Twinhead
420-490nm
Plus Light Head
385-490nm
Violet Light Head
385-440nm
White Light Head
5000- 6500K
fPad Head
620-660nm
Tip Size (mm)
Standard Head
10
Plus Light Head
12
Light Intensity
LED Light Head
4000 (Plasma Mode); 2000 (Pulse Mode)
Tacking Tip