Product Safety Warnings and Cautions 

Prior to installation and activation of the light, please read these instructions carefully and follow safety 
procedures as well as the specific safety comments and recommendations presented throughout the 
instructions for use. The manufacturer accepts no liability for any damage resulting from the improper use 

of this unit and/or for any purpose other than those covered by these instructions.

 

1.

 

          CAUTION! FUSION-DOE produces high intensity light. Prolonged exposure to the skin, oral mucosa, or 
pulp can result in irritation or injury.   

-

 

  Do not place light in direct contact with unprotected gingival, oral mucous membrane or skin.   

-

 

  Do not place light directly on the same tooth segment for more than 6 seconds in Plasma Power Mode. 

-

 

  If long exposure cannot be avoided, direct a stream of air over the tooth surface from an air syringe or 

high-speed vacuum tip, adjust to Pulse Power Mode or adjust curing distance. 

2.

 

          CAUTION! Do not look directly at emitted light. Do not use FUSION-DOE without suitable eye protection 
for the operator, assistant, and patient. Use Light Shields, Goggles, and/or Curing Caps. Exposure must be 
restricted to the area of the oral cavity for which the clinical treatment is intended.   

3.

 

          CAUTION! FUSION-DOE 

must be used with “Single Use” Barrier Sleeves to avoid cross contamination.

 

Replace Barrier Sleeve after each patient treatment is completed. This sleeve also prevents composite/residue 
material from adhering to the surface of the lens and degrading the curing performance.   

4.

 

Products 

marked “Single Use” on the labeling are intended for single use only. Discard properly after each use. 

 

5.

 

DO NOT operate the unit in the proximity of flammable mixtures. Use in well-ventilated area. 

6.

 

Keep solvents, flammable liquids, and sources of intense heat away from FUSION-DOE as they may damage 
internal and external parts, the seals, or the buttons on the control panel.   

7.

 

DO NOT autoclave or immerse the unit in any kind of ultrasonic bath, disinfectant, cleaning solution or any kind 
of liquid. Do not spray liquids directly onto the light and prevent liquids from entering the device open area such 
as Keypad, Charging Stand, and its connectors as this may damage or short-circuit the device and can cause 
potential hazard for operator and/or patient. 

Start Up 

Assemble Handpiece   

Upon receiving your kit, 
remove all components from 
the package and examine 
them. If damage is apparent, please contact your authorized DentLight distributor from whom your light was purchased.   
Attach the Battery Assembly to the Main Control by screwing the Battery Assembly clockwise into the Main Control. The 
silver FUSION head is attached to the Main control for composite curing. Adjust the 360º swivel Light Head to a 
comfortable curing position and ensure good connection with the Main Control. The Light Head quick connects with the 
Main Control with easy twist Push/Pull and can easily change with the DOE violet light head for fluorescent imaging. Light 
is now ready for use.   

Operation and Setting

 

The handpiece uses advanced Metal Keypad and LCD interface. ON/OFF Key starts and stops the light. TM/PW 
Key sets the Operation Time and Power Mode. The light defaults in Pulse Power Mode (green backlit LCD and flash 
MODE symbol) with 5 second timer. Press TM/PW Key to select a Digital Timer of your choice. Press and hold 
TM/PW Key for 2 seconds and release to change power to Plasma Power Mode (white backlit LCD with steady 
MODE symbol) and vice versa.     
Plasma Power Mode allows the fastest curing with plasma-like energy dosage. Pulse Power Mode maintains the 
peak power and generates the minimum heat by lowering the average power in half with a 50/50 on and off duty 

 

 

 

Summary of Contents for FUSION-DOE

Page 1: ...LED Curing Exam Combo Light System Powered by Patented Focus Beam Technology Setting New Standards through Innovation FUSION DOE INSTRUCTIONS FOR USE ...

Page 2: ...rrier Sleeves Stand Stick Pad and Instructions for Use Quantity Description 1 FUSION Light Head 1 DOE Light Head 1 Main Control 1 Battery Assembly 3 Curing Caps 3mm 9mm 13mm each 1 Barrier Sleeves 1 pack of 100 1 Orange Fluorescent Shield 1 Green Fluorescent Shield 1 Orange Goggle 1 Green Goggle 1 Power Adapter 1 Light shield 1 Charging Stand 1 Stand Stick Pad 1 Instructions for Use The handpiece ...

Page 3: ...lvents flammable liquids and sources of intense heat away from FUSION DOE as they may damage internal and external parts the seals or the buttons on the control panel 7 DO NOT autoclave or immerse the unit in any kind of ultrasonic bath disinfectant cleaning solution or any kind of liquid Do not spray liquids directly onto the light and prevent liquids from entering the device open area such as Ke...

Page 4: ...l the same if well cured For general guidance use recommended curing times below For reduced heat or shrinkage use Pulse Power Mode and replace 3sec with 5sec curing times below Recommended Cure Times seconds Plasma Power Mode Manufacturer 3M Ivoclar Dentsply Kerr Resin Brand Filtek Supreme Plus Heliomolar Tetric EvoCeram TPH3 XRV Herculite Color Shade A3E C6D A3 A2 A3 DY A3 YB Cure Times Per laye...

Page 5: ...y exams but also for preps to clean up all residue caries and infections Battery Care Keep the device on the Charging Stand when not in use Connect DC Power Plug to Plug Receptor on the Charging Base Peel off Stand Stick Pad and secure Charging Stand on a table counter for easy one handed pick up Alternatively connect DC Power Plug directly to the back end of the Handpiece to charge and power on t...

Page 6: ...Add ons FUSION DOE is the most advanced and versatile LED light system allowing users to perform beyond normal and composite curing and exams You can implement additional advanced curing and therapeutic procedures by twisting and pulling out its modular light head and switch it for custom illumination Additional optional light heads are Plus Head providing larger 12mm curing tip with broader wavel...

Page 7: ...ce Lifetime Lifetime of the device is approximately 150 000 curing cycles based on battery life or typically 5 service years in normal office use Operating environment Ambient temperature 50ºF to 86ºF 10ºC to 30ºC Relative humidity 10 80 non condensing Atmospheric pressure 0 5 atm to 1 0 atm 500hPa 1060hPa Storage environment Ambient temperature 14ºF to 104ºF 10ºC to 40ºC Relative humidity 10 80 n...

Page 8: ...maged 3 Replace battery Light generates too much heat Patient feeling discomfort 1 Cure 3 seconds at plasma power or 5 seconds at pulse mode per layer 2 Direct an air stream from an air syringe or high speed vacuum tip 3 Adjust to Pulse Power Mode 4 Adjust curing distance Battery will not charge 1 Broken pins in Charging Stand 2 Electronic defect in Battery Assembly 1 Bypass replace or service Cha...