Product Safety Warnings and Cautions
Prior to installation and activation of the light, please read these instructions carefully and follow safety
procedures as well as the specific safety comments and recommendations presented throughout the
instructions for use. The manufacturer accepts no liability for any damage resulting from the improper use
of this unit and/or for any purpose other than those covered by these instructions.
1.
CAUTION! FUSION-DOE produces high intensity light. Prolonged exposure to the skin, oral mucosa, or
pulp can result in irritation or injury.
-
Do not place light in direct contact with unprotected gingival, oral mucous membrane or skin.
-
Do not place light directly on the same tooth segment for more than 6 seconds in Plasma Power Mode.
-
If long exposure cannot be avoided, direct a stream of air over the tooth surface from an air syringe or
high-speed vacuum tip, adjust to Pulse Power Mode or adjust curing distance.
2.
CAUTION! Do not look directly at emitted light. Do not use FUSION-DOE without suitable eye protection
for the operator, assistant, and patient. Use Light Shields, Goggles, and/or Curing Caps. Exposure must be
restricted to the area of the oral cavity for which the clinical treatment is intended.
3.
CAUTION! FUSION-DOE
must be used with “Single Use” Barrier Sleeves to avoid cross contamination.
Replace Barrier Sleeve after each patient treatment is completed. This sleeve also prevents composite/residue
material from adhering to the surface of the lens and degrading the curing performance.
4.
Products
marked “Single Use” on the labeling are intended for single use only. Discard properly after each use.
5.
DO NOT operate the unit in the proximity of flammable mixtures. Use in well-ventilated area.
6.
Keep solvents, flammable liquids, and sources of intense heat away from FUSION-DOE as they may damage
internal and external parts, the seals, or the buttons on the control panel.
7.
DO NOT autoclave or immerse the unit in any kind of ultrasonic bath, disinfectant, cleaning solution or any kind
of liquid. Do not spray liquids directly onto the light and prevent liquids from entering the device open area such
as Keypad, Charging Stand, and its connectors as this may damage or short-circuit the device and can cause
potential hazard for operator and/or patient.
Start Up
Assemble Handpiece
Upon receiving your kit,
remove all components from
the package and examine
them. If damage is apparent, please contact your authorized DentLight distributor from whom your light was purchased.
Attach the Battery Assembly to the Main Control by screwing the Battery Assembly clockwise into the Main Control. The
silver FUSION head is attached to the Main control for composite curing. Adjust the 360º swivel Light Head to a
comfortable curing position and ensure good connection with the Main Control. The Light Head quick connects with the
Main Control with easy twist Push/Pull and can easily change with the DOE violet light head for fluorescent imaging. Light
is now ready for use.
Operation and Setting
The handpiece uses advanced Metal Keypad and LCD interface. ON/OFF Key starts and stops the light. TM/PW
Key sets the Operation Time and Power Mode. The light defaults in Pulse Power Mode (green backlit LCD and flash
MODE symbol) with 5 second timer. Press TM/PW Key to select a Digital Timer of your choice. Press and hold
TM/PW Key for 2 seconds and release to change power to Plasma Power Mode (white backlit LCD with steady
MODE symbol) and vice versa.
Plasma Power Mode allows the fastest curing with plasma-like energy dosage. Pulse Power Mode maintains the
peak power and generates the minimum heat by lowering the average power in half with a 50/50 on and off duty