15
www.DentalEZ.com
866-DTE-INFO
15
DentalEZ
®
Equipment
PN: 2717-256E
Section III Installation
Air Glide
(Optional)
NOTE:
If the air glide option was ordered,
the air regulator was attached to the base
chassis prior to shipment of the chair.
1. Place the NuSimplicity Chair on a smooth, hard
and level (vinyl, tile or terrazzo type)
fl
oor.
2. Connect the air glide tubing to a regulated air
supply source (usually 552 kPa [80 PSI] on
most dental units)
(Figure 15)
.
3. The air glide is pre-set at the factory to
138 kPa (20 PSI). Because of additional
optional equipment, it may be necessary to
increase the pressure. Slowly increase the
air pressure by adjusting the air regulator
on the base chassis until the chair begins to
“bounce”. This is an indication that the air
pillow is over-pressurized and can not seal to
the
fl
oor. Slowly decrease the pressure until
the chair stops bouncing.
4. Actuate the air regulator toggle switch
(Figure 16)
at the rear of the chair on the base
cantilever section cover.
Air Regulator
Air Glide
Tubing
Figure 15. Air glide tubing connected to air regulator
5. Check for proper air glide operation.
6. Make sure the doctor and sta
ff
receive proper
air glide feature operation instructions (see
Air
Glide
in
Section IV
Operation
of this manual).
Air Regulator
Toggle Switch
Figure 16. Air regulator toggle switch
For proper air glide operation, the
fl
oor MUST
BE LEVEL and have a smooth, hard and non-
obstructive surface.
NOTICE
Summary of Contents for NuSimplicity
Page 1: ...NuSimplicityTM Chair User Manual ...
Page 3: ...www DentalEZ com 866 DTE INFO ii NuSimplicityTM Chair PN 2717 256E ...
Page 29: ...www DentalEZ com 866 DTE INFO 26 NuSimplicityTM Chair PN 2717 256E ...
Page 33: ...www DentalEZ com 866 DTE INFO 30 NuSimplicityTM Chair PN 2717 256E ...
Page 35: ...www DentalEZ com 866 DTE INFO 32 NuSimplicityTM Chair PN 2717 256E ...