13
Activating Alarm Set-points for the Second Input Channel/
(
relative humidity in HD206-1
and HD206-2 models and External Temperature in HD207 and HD207-1 models).
The instrument indicates that humidity and external temperature alarm set-points (as set at step
P08) have been exceeded and lights up the
symbol on display in correspondence with
ALRM indication. To enable this function, use T3 (
) and T4 (
) keys to select ON, then
press ENTER to confirm (T1).
P09
off
[ ]
[ ]
1
2
Note: the alarm indication remains even if, after going out of range, the measurement re-
turns to set limits. To reset previous alarm indications, set OFF and press ENTER (T1)
to confirm.
Setting the Instrument in a Default Configuration
.
“deF” indication appears: press ENTER (T1) to confirm or T4 to jump to P0 without making
any change.
The default configuration allows to set the following parameters:
•
Sampling time: 5 sec
•
Manual Start
•
Endless Loop Memory: OFF
•
Alarm set-points for the first channel: OFF