HD402TR
- 9 -
V1.0
Command
Description
ART1
Reads the threshold 1 value
AWT2snnnn Sets the threshold 2 value to snnnn (“s” is the sign of the value)
(*)
Default: pressure value corresponding to 70% of the full scale
ART2
Reads the threshold 2 value
AWHnnnn
Sets the hysteresis value to nnnn
(*)
Default: pressure value corresponding to 10% of the full scale
ARH
Reads the hysteresis value
AWD1nnn
Sets the alarm activation delay to nnn seconds (0… 600 s)
Default: 0
ARD1
Reads the alarm activation delay value
AWD2nnn
Sets the alarm deactivation delay to nnn seconds (0… 600 s)
Default: 0
ARD2
Reads the alarm deactivation delay value
(*)
The thresholds and hysteresis values are considered in the unit of measurement set
in the instrument. The value must be written without the decimal point, even if it is not
an integer value (e.g., to set thresholds 1 to +1.500, write AWT1+1500). Leading zeros
can be omitted (e.g., to set thresholds 1 to +0.050, write AWT1+50).
To exit the configuration mode after sending the CAL START command, send the
CAL END command (the instrument automatically exits the configuration mode after 5
minutes from the last command sent).
C
ONFIGURATION VIA THE INTERNAL BUTTONS
The configuration set with the internal buttons is used by the instrument only if the
dip switch 1 is ON.
The upper button B1 allows scrolling the available operating parameters, while the
lower button B2 allows changing the setting of the selected parameter.
The function of a button depends on whether it is short or long pressed. To short
press a button, hold it down for about 1 second, until the power LED turns off. To long
press a button, hold it down for at least 3 seconds, until the power LED turns off and then
turns on again.
Long press the upper button B1 to enter the menu. Inside the menu, long press B1 to
scroll the available parameters. Below is the sequence of the operating parameters (in
brackets the indication that appears on the display) with the possible settings for non-
numerical parameters:
•
Relay operating mode (SECU):
o
Negative (NEG): NC contact is closed if no alarm, NO contact is closed if in alarm
o
Positive (POS): NO contact is closed if no alarm, NC contact is closed if in alarm
•
Buzzer activation (BEEP):
o
OFF: buzzer disabled
o
ON: buzzer enabled
•
Alarm activation (ALAR):
o
OFF: alarm disabled
o
ON: alarm enabled