![Delta OHM HD40.2 Manual Download Page 4](http://html1.mh-extra.com/html/delta-ohm/hd40-2/hd40-2_manual_2478320004.webp)
HD402TR
- 4 -
V1.0
2
TECHNICAL CHARACTERISTICS
Sensor
Piezoresistive, High stability
Measuring range
From ±250 Pa to ±200 kPa both relative and differential
Refer to table 1
Resolution
Refer to table 2
Accuracy @ 25 °C
± 1.5% f.s. nominal for HD402TR1L
± 0.75% f.s. nominal for HD402TR2L
± 1% f.s. nominal for HD402TR3L, HD402TR4L and HD402TR5L
Accuracy @ 0…50 °C
± 3% f.s. nominal for HD402TR1L
± 1% f.s. nominal for HD402TR2L, HD402TR3L, HD402TR4L and
HD402TR5L
Long term stability (1000 h)
@ 25 °C
± 0.5% f.s. nominal for HD402TR1L and HD402TR2L
± 0.35% f.s. nominal for HD402TR3L
± 0.25% f.s. nominal for HD402TR4L and HD402TR5L
Alarm
Front LED, internal buzzer, relay switch
Output
SPDT Relay switch
3 A/250 Vac
3 A/30 Vdc resistive load
Connection to PC
RS232 serial port
Can be connect to a USB port by using the optional CP27 adapter
Configuration
Settable unit of measurement, thresholds, hysteresis, delay and alarm
operation mode
Auto-zero
Automatic for HD402TR1L, manual for the other models
Response time
0.5 seconds for the display updating
Overpressure limit
50 kPa for HD402TR1L, HD402TR2L and HD402TR3L
200 kPa for HD402TR4L
400 kPa for HD402TR5L
Compatible media
Only air and non-aggressive dry gases
Power supply
24 Vac
±
10% or 15…36 Vdc
Absorption
< 1 W @ 24 Vdc
Pressure connection
Ø 6.2 mm for tubes with internal Ø 5…6 mm
Electrical connections
Screw terminal block, max 1.5 mm
2
, PG9 cable gland for the input cable
Operating conditions
-10…+60 °C / 0…95% RH
Storage temperature
-20…+70 °C
Housing dimensions
80 x 84 x 44 mm
Protection degree
IP65
TAB. 1: full scale values and units of measurement
Model
Pa
kPa
mbar
mmH
2
O
inchH
2
O
mmHg
PSI
HD402TR1L
250
--- 2.5 25 1 --- ---
HD402TR2L
1000 ---
10 100 4
--- ---
HD402TR3L
--- 10 100 --- --- 50 1.5
HD402TR4L
--- 100 1000 --- --- 500 15
HD402TR5L
--- 200 2000 --- --- 1000 30
Summary of Contents for HD40.2