HD32.3
- 20 -
V1.4
4.
WORKING
Before switching the instrument on, connect the SICRAM probes to the inlets: 8 pin male
connector DIN 45326 that
is on the bottom of the instrument.
NOTE:
The probes must be connected to the instrument off. If you connect a new probe to the
instrument already switched on, it isn't recognized, so it's necessary to switch the instrument
off and switch it on again.
If you disconnect a probe with the instrument switched on, you have an acoustic alarm (a beep per
second) and you can visualize it on the display in correspondence of a disconnected physical size,
the “LOST” message will be visualized.
If you insert more probes of the same type, it will be only consider the first recognized probe:
the scansion of the probes, for their recognition, happens from the inlet 1 up the inlet 3.
When you switch the instrument on, on the display appears the following writing for at least 10 sec-
onds:
HD32.3 WBGT - PMV
Firm.Ver.=01.00
In addition to Delta
Ohm
logo are
indicated the
instrument code and the firmware version.
After connecting the probes, switch the instrument on: the display, after 10 seconds, will appear in
the measurement visualization mode that refers to WBGT and PMV indexes.
WBGT Index
2008/11/28 08:00:00
Tn 15.6 °C
Tg 20.2 °C
T 20.2 °C
WBGT(in) 17.0 °C
WBGT(out) 17.0 °C
Tn:
wet bulb temperature with natural ventilation
Tg:
globe thermometer temperature, detected by the globe thermometer probe
T:
ambient temperature, detected by Pt100 probe
WBGT (in):
WBGT index calculated in absence of solar irradiation
WBGT (out):
WBGT index calculated in presence of solar irradiation
Summary of Contents for HD32.3
Page 4: ...HD32 3 4 V1 4 HD32 3 WBGT PMV PPD indices...
Page 52: ...HD32 3 52 V1 4 12 PRINTING OF THE MEASUREMENT REPORT Norm ISO 7243 Norm ISO 7243...
Page 53: ...HD32 3 53 V1 4 Warm environments Norm ISO 7243...
Page 54: ...HD32 3 54 V1 4 Warm environments Norm ISO 7243 Norm ISO 7243...
Page 55: ...HD32 3 55 V1 4 Warm environments Norm ISO 7243...
Page 56: ...HD32 3 56 V1 4 Warm environments Norm ISO 7243...
Page 57: ...HD32 3 57 V1 4 Warm environments Norm ISO 7243...
Page 58: ...HD32 3 58 V1 4 Moderate environments Norm ISO 7730 Norm ISO 7730...
Page 59: ...HD32 3 59 V1 4 Moderate environments Norm ISO 7730...
Page 60: ...HD32 3 60 V1 4 Moderate environments Norm ISO 7730...
Page 61: ...HD32 3 61 V1 4 Moderate environments Norm ISO 7730...
Page 62: ...HD32 3 62 V1 4 Moderate environments Norm ISO 7730...
Page 63: ...HD32 3 63 V1 4 Moderate environments Norm ISO 7730...
Page 64: ...HD32 3 64 V1 4 Moderate environments Norm ISO 7730...
Page 71: ......