HD2307
- 16 -
V1.7
I N STRUM EN T
SI GN ALS
AN D
FAULTS
The follow ing t able list s all error indicat ions and inform at ion displayed by t he
inst rum ent and supplied t o t he user in different operat ing sit uat ions:
D ispla y in dica t ion s
Ex pla n a t ion
> > > CAL_ M OD E
Calibrat ion m ode
RTD _ PRBE_ TYPE
Type of pr obe connect ed
1 ST_ M EAS UP D OW N
Cor r ect t he first point using t he arr ows
/
2 N D _ M EAS UP D OW N
Cor r ect t he second point using t he arrows
/
BATT TOO LOW
CH N G N OW
I ndication of insufficient batt ery charge appearing on turning on.
The inst rum ent issues a long beep and turns off. Replace the
batt eries.
CAL
LOST
Program err or: it appears aft er t urning on for a few seconds.
Cont act t he inst rum ent 's supplier.
CAL FACT
Fact or y calibrat ion
CAL USER
User calibrat ion
ERR
This appears if the probe has already been det ect ed by the
instrum ent, but is disconnect ed. At the sam e tim e an
interm ittent beep is issued.
FUN C CLRD
Max, m in and average v alues cleared
N EW PROBE D ET
This m essage appears when a probe is inserted into a functioning
instrum ent. Turn the instrum ent off and then back on again.
N O_ PRBE_ SER_ N UM
The connect ed probe's serial num ber is absent
OVER or UN D R
Measurem ent overflow : indicat es t hat t he probe is m easuring
a value ex ceeding t he m easuring range.
PLS_ EX I T > > > FU N C
RES_ FOR_ FACT ON LY
Please exit using ESC > > > funct ion reserv ed t o fact ory
calibrat ion
PRBE_ SER # # # # # # # #
Serial num ber # # # # # # # # of t he connect ed probe
PROB
ERR
A probe wit h SI CRAM m odule has been insert ed when not
adm issible for t hat specific inst rum ent .
PROB COM M LOST
This appears if t he pr obe has already been det ect ed by t he
inst rum ent , but is disconnect ed. At t he sam e t im e an
int erm it t ent beep is issued.
SEL M EAS 1 / 2
Select m easur em ent 1 or 2
SYS
ERR
#
I nst rum ent m anagem ent program er ror. Cont act t he
inst rum ent 's supplier and com m unicat e t he num eric code #
report ed by t he display.
I N STRUM EN T
STORAGE
I nst rum ent st orage condit ions:
•
Tem perat ure: - 25…+ 65° C.
•
Hum idit y: less t han 90% RH wit hout condensat ion.
•
Do not st ore t he inst rum ent in places where:
•
hum idit y is high;
•
t he inst rum ent m ay be exposed t o direct sunlight ;
•
t he inst rum ent m ay be exposed t o a source of high t em perat ure;
•
t he inst rum ent m ay be exposed t o st rong vibrat ions;
•
t he inst rum ent m ay be exposed t o st eam , salt or any corrosive gas.
The inst rum ent case is m ade of ABS plast ic: do not use any incom pat ible solvent for
cleaning.