HD2013.3 

-   10   -

 

V1.7

 

5

 

PROPRIETARY PROTOCOL 

To use the proprietary protocol, it is necessary to connect the instrument to the PC via 
a  RS485/USB  (e.g.  RS51K)  or  RS485/RS232  converter  and  use  a  standard  serial 
communication  program.  In  the  serial  communication  program,  set  the  COM  port 
number  to  which  the  instrument  is  connected  and  the  communication  parameters  as 
follows: 

 

If  the  MODBUS-RTU  protocol  is  set  as  the  operating  protocol  in  the  instrument 
(default), set the Baud Rate 57600 and the parameters 8N2 in the serial commu-
nication  program,  then  power  cycle  the  instrument  and  send  the  command 

@

 

within 10 seconds from the instrument power on. 

 

If the proprietary protocol is already set as the operating protocol in the instru-
ment,  it  is  possible  to  operate  with  Baud  Rate  57600  and  parameters  8N2  by 
sending  the  command 

@

  within  10  seconds  from  the  instrument  power  on,  or 

you  can  let  the  10  seconds  pass  without  sending  the  command 

@

  and  operate 

with the communication parameters set in the instrument (default 19200, 8E1). 

To  change  the  instrument  configuration,  the  serial  command 

CAL  USER  ON

  must  be 

sent first (the instrument replies with USER CAL MODE ON). The command CAL USER ON 
is automatically disabled after a few minutes of inactivity. If the settings should be only 
read, the command CAL USER ON is not required. 

Below is the list of the serial commands. 

Instrument information: 

Command 

Reply

 

Description 

 

G0 

Model 

Instrument model 

G1 

&

Revision

Instrument hardware revision 

G2 

SN=nnnnnnnn 

Instrument serial number 

G3 

Firm.Ver.=x.y 

Instrument firmware revision 

G4 

Firm.Date=yyyy/mm/dd 

Date of firmware revision 

GC 

Fact.Calib.Date= yyyy/mm/dd 

Date of factory calibration 

Protocol:

 

Command 

Reply

 

Description 

 

&| 

Keeps the proprietary protocol operational even after 10 
seconds from instrument power on. It must be sent within 
10 seconds from instrument power on. 

CPn 

&| 

Sets the operating protocol: 

 

Proprietary if n=0 

 

MODBUS-RTU if n=1 

Default 

: MODBUS-RTU (n=1) 

RP 

& n| 

Reads the operating protocol set in the instrument. 

SM 

&| 

Activates the MODBUS-RTU protocol immediately. 

CMAn 

&| 

Sets the instrument address for the MODBUS-RTU pro-
tocol to n. 
The address should range within 1 and 247. 

Default 

: 1 

RMA 

& n| 

Reads the instrument address for the MODBUS-RTU pro-
tocol. 

Note

: after sending the CP1  command, the instrument  remains with the proprietary protocol. 

Send the command SM to activate the MODBUS-RTU protocol immediately, or power cycle the 

Summary of Contents for HD2013.3

Page 1: ...Operating manual Precipitation detector HD2013 3 www deltaohm com English Keep for future reference...

Page 2: ...INTRODUCTION 3 2 TECHNICAL SPECIFICATIONS 4 3 INSTALLATION 6 3 1 ELECTRICAL CONNECTIONS 6 4 CONFIGURATION AND MEASUREMENT 9 5 PROPRIETARY PROTOCOL 10 6 MODBUS RTU PROTOCOL 17 7 MAINTENANCE 21 8 SAFETY...

Page 3: ...instrument external circular dome acts as a windshield for the sensor preventing false indications An auto calibration algorithm compensates for variations in the response due to the dirt that can ac...

Page 4: ...ensor Temperature 0 1 C Output RS485 not isolated with MODBUS RTU or proprietary protocol Analog 0 4 20 mA RLmax 500 or on request 0 10 V RLmin 10 k Voltage free normally open contact max 200 mA 30 Vd...

Page 5: ...HD2013 3 5 V1 7 Fig 2 1 dimensions mm Bird spikes optional Bracket...

Page 6: ...is free from any object that could obstruct the detection of the precipitation Keep the sensor clean 3 1 ELECTRICAL CONNECTIONS The instrument has a 8 pole circular connector and uses the CP2013 2 opt...

Page 7: ...ing the proprietary protocol or the MODBUS RTU protocol Fig 3 3 RS485 connection Current analog output By default the current analog output is 4 20 mA with 4 mA sensor sensitive surface completely wet...

Page 8: ...rection of the output so that when the sensor sensi tive surface is completely wet the output is 10 V In case of measurement error the output goes to 11 V Fig 3 5 voltage analog output connection Cont...

Page 9: ...ddress 5 The indication of precipitation in progress ceases when the percentage of dry sensitive surface of the sensor rises above the threshold plus a configurable hysteresis value default 2 and afte...

Page 10: ...the instrument replies with USER CAL MODE ON The command CAL USER ON is automatically disabled after a few minutes of inactivity If the settings should be only read the command CAL USER ON is not req...

Page 11: ...ates protocol Waiting 3 5 characters if n 1 respects protocol Default Immediate reception n 0 RMW n Reads the setting of waiting time after transmission with MODBUS RTU protocol Precipitation detectio...

Page 12: ...RSH n Reads the setting of the hysteresis of the reference tem perature to discriminate between rainfall and snowfall CRADn Sets the delay in indicating the beginning of a precipita tion both rainy a...

Page 13: ...ered a rainfall if the measured temperature is higher than the reference temperature and a snowfall if the measured tempera ture is lower than the reference temperature Default Output activated in cas...

Page 14: ...CENTAGE Prints the list of the measurement information sent by the instrument RAIN_RAW number of 12 bit ADC counts proportional to the percentage of dry sensitive surface of the sensor RAIN_LEVEL perc...

Page 15: ...if the sensor sensitive surface is completely dry CRBSn Enable disable the precipitation sensor automatic cali bration at regular intervals Disable if n 0 Enable if n 1 Default Automatic calibration...

Page 16: ...her hand the greater the possibil ity that a very light precipitation is not detected or causes an oscillation in indicating the presence absence of pre cipitation The T setting is therefore a comprom...

Page 17: ...owfall Bit Input Registers Address Description Format 0 Instrument model identification 16 bit Integer 1 Instrument hardware revision 16 bit Integer 2 3 Instrument serial number 16 bit Integer 4 Instr...

Page 18: ...Format 0 RS485 Baud Rate 0 1200 1 2400 2 4800 3 9600 4 19200 default 5 38400 6 57600 7 115200 16 bit Integer 1 RS485 parity and stop bits 0 8N1 1 8N2 2 8E1 default 3 8E2 4 8O1 5 8O2 N no parity E eve...

Page 19: ...activation mode 0 Output disabled 1 Output activated in case of rainfall default 2 Output activated in case of snowfall Note the precipitation is considered a rainfall if the measured temperature is...

Page 20: ...HD2013 3 20 V1 7 Address Description Format 21 Sensitivity of the precipitation detection 10 1000 default 100 The greater the value the greater the sensitivity 16 bit Integer...

Page 21: ...ent control and laboratory use and has left the factory in perfect safety tech nical conditions The instrument proper operation and operating safety can be ensured only if all standard safety measures...

Page 22: ...s to be ordered separately HD2013 2D Bird spikes 6 spikes 3 mm height 60 mm To be requested when ordering the instrument CP2013 2 5 Cable with 8 pole watertight connector on one end free wires on the...

Page 23: ......

Page 24: ...rranty included in the manual of the product For any dispute the competent court is the Court of Padua The Italian law and the Convention on Contracts for the International Sales of Goods apply TECHNI...

Reviews: