48 / 53
5.12
T
EST
P
ARAMETERS
P80:
Current output test (mA)
Going to this parameter, the actual current output (corresponding to the measured process value) will be displayed. By pressing ENTER
E
the
(now blinking) current value can be set for any value between 3.9 and 20.5 mA. The current output has to show the same value which can be
checked by an ampere meter, according to the description under 4.4. Press ENTER
E
to quit test mode and return the parameter address
P81: - - - a Relay test
The actual state of the relay can be seen on the display (code according to the table below and symbol on the screen). Test the relay by pressing
UP and DOWN while observing change of the symbol and the code or listening to the ticking of the relay or checking on-off resistance by a
siutable resistance meter.
a
Relay state
0
De-energised
1
Energised
P97: b:a.aa Software code
a.aa:
Number of the software version
b:
Code of the special version