26 / 53
P01: - ba
Measurement Mode
–
Bargraph
Parameter value „a” will determine the basic measurement value that will be displayed and proportional with t
he current output. Depending on the
value of “a” process values as listed in the 3d colu
mn can also be displayed by pressing NEXT . For return to the display of the basic value the
ENTER
E
key should be pressed.
a
Measurement Mode
Display symbol
Displayed values
Attention: mind the sequence!
When programming this
parameter the
right value “a”
will be blinking first.
0
Distance
DIST
Distance
1
Level
LEV
Level, Distance
2
Level in percentage
LEV%
Level%, Level, Distance
3
Volume
VOL
Volume, Level, Distance
4
Volume in percentage
VOL%
Volume%, Volume, Level, Distance
5
Flow
FLOW
Flow, TOT1, TOT2, Level, Distance
Parameter value “b” will determine that the height of the Bargraph will
be proportional to the current output or to the Echo strength.
b
Bargraph indication
0
Echo strength
1
Current output
FACTORY DEFAULT: 11
P02: - cba Calculation units
a
Temperature
Attention: mind the sequence!
When programming this parameter the right
value “a” will be blinking f
irst.
0
°C
1
°F
This table is interpreted according to
P00(c)
,
P01(a)
and
P02(c)
and is irrelevant in case of percentage measurement (
P01(a)= 2
or
4
)
b
Volume
Weight (set also P32)
Volume flow
Metric
US
Metric
US
Metric
US
0
m
3
ft
3
-
lb (pound)
m
3
/time
ft
3
/time
1
liter
gallons
tons
tons
liter/time
gallons/time
c
Time
0
Sec
1
Min
2
Hour
3
Day
FACTORY DEFAULT: 000