4. SPECIFICATIONS
Range:
“C” Range
+1472 °F to +3092 °F (+800 °C to +1700 °C)
“D” Range
+662 °F to +3092 °F (+350 °C to +1700 °C)
“F” Range
Dual Wavelength
Single Wavelength
+1472 °F to +3092 °F (+800 °C to +1700 °C)
+662 °F to +3092 °F (+350 °C to +1700 °C)
Accuracy
0.5 % of reading or 36
F (±20
C), whichever is greater
Repeatability
0.1 % of full-scale span
Warm-Up Time
60 seconds
Readout
High Contrast LCD Display
Output Signal:
4-20 mADC, 2-wire loop-powered
Hazardous Location Rating
II 2 G Ex d IIC T6 (When installed per section 2.3.1 Remote Sensor
Mounting — Option RS.)
IP65
Fiber Optic Cable:
Flexible PTFE over SS armor
10 feet (3 meters) standard length
Minimum Bend Radius
3 inches (7.6 cm)
Hot Lens Mounting
3” ANSI is standard on a flanged nozzle steam jacketed.
Other types and sizes available
Ports:
Steam Jacket in and out
50 psi to 100 psi (344 kPa to 689 kPa) for steam to heat the lens
assembly
Nitrogen Purge
60 SCFH (28 L/min) typical
Housing Mounting
Vertical surface or 2-20 pipe stand with vertical or horizontal
orientation
Housing Conduit Connection
¾” NPT
Ambient Temperature Limits:
Fiber Optics & Lens
-20
F to +400
F (-29
C to +204
C)
Fiber Optics Adapter
-4
F to +185
F (-20
C to 85
C)
Electronics
-4
F to +158
F (-20
C to +70
C)
Maximum Process Pressure
150 psi (1034 kPa)
585 Fortson Street
Shreveport, LA 71107
www.claustemp.com