optical sight path and performing any necessary sight glass or lens cleaning and / or nozzle cleanout. Refer to
section 2.5 Optical Alignment for details.
3.5 Calibration
Calibration should be checked any time an anomaly is suspected or annually. The best way to check the
calibration or to recalibrate the instrument is with a CLAUSTEMP
®
Model HIP Handheld Pyrometer. See
www.claustemp.com
.
To calibrate the HIR Pyrometer:
1.
Measure reactor temperature using a CLAUSTEMP
®
model HIP handheld pyrometer or other means.
2.
Ensure Model HIR has a clear sight path into the reactor.
3.
Press and hold the NEXT button while simultaneously pressing the UP arrow button. Then release both
buttons.
4.
Press the NEXT button to display the CAL TEMP parameter.
5.
Unlock the calibration feature by pressing and holding the NEXT button while simultaneously, pressing
the DOWN arrow button. Then release both buttons.
6.
Use the UP and Down arrow buttons to set the known reactor temperature.
7.
Press the SAVE button. Calibration is complete.
00-HIR05
Fiber Optic Cable
3” Minimum Bend Radius
Figure 12 - Dimensions – Steam Jacketed Lens Assembly
Units: [mm]
inches