Trigno Spring Contact Adapter User’s Guide
12
Trigno systems are shipped with all sensors appropriately paired.
Sensor pairing is typically needed in the following situations: a)
if sensors are being replaced within the network group, b) when
the communication frequency sets are changed, and c) after a
firmware upgrade is performed.
Upon pairing sensors, EMGworks will present the option to en-
ter a “Delsys Factory Calibration” sequence or to use the “Auto-
Detected Calibration”. Select the “Auto-Detected Calibration”
option unless the sensor is specifically supplied with a calibration
key (this is not common). Note that if the sensor is being paired
for the first time with the base station, the choice will read “Use
Default Calibration” rather than “Use Auto-detected Calibra-
tion”.
Figure 5:
The calibration can be auto-detected, or manually re
-
trieve..
Smart Sensor Features
After pairing, the association of sensors to the Trigno System is
retained for all future uses. Any configuration in EMGworks can
be made to reflect the last paired set of sensors by clicking the
“Refresh Smart Sensors” button in the “Add Sensors” pane in
EMGworks. When data collection starts, the software will verify
that the sensors currently communicating match those used in
the configuration. If there is a mismatch, cancel the recording
and repair the sensors.
The Spring Contact Adapter will appear with a unique icon
shown below in EMGworks once it has been properly paired
and identified.